Quantifying the effects of positional uncertainties and estimating margins for Gamma-Knife ® fractionated radiosurgery of large brain metastases
Besides conventional single fraction stereotactic radiosurgery (SRS), the newer Gamma Knife® (Elekta Instruments, AB, Sweden) (GK) models, Perfexion™ and Icon™, also allow the option of hypofractionated SRS, to widen the treatment range, reduce toxicity and provide flexibility. With fractionation, s...
Saved in:
Published in: | Journal of radiosurgery and SBRT 2017-01, Vol.4 (4), p.275-287 |
---|---|
Main Authors: | , , , |
Format: | Article |
Language: | English |
Subjects: | |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Besides conventional single fraction stereotactic radiosurgery (SRS), the newer Gamma Knife® (Elekta Instruments, AB, Sweden) (GK) models, Perfexion™ and Icon™, also allow the option of hypofractionated SRS, to widen the treatment range, reduce toxicity and provide flexibility. With fractionation, set-up uncertainties are introduced to the treatment. The question of what margin is required to cover set-up uncertainties arises. In this work, the characteristics of positional uncertainties and margins required to ensure coverage are evaluated. An alternative strategy on how to deal with set-up errors is suggested.
Five large single metastases have been considered to simulate the dosimetric effects of set-up uncertainties. Since random error would not create a representative average for three or five fractions a "worst case scenario" with maximal displacement in different directions has been evaluated. Displacements of 0.5, 1.0, 2.0, 3.0 and 4.0 mm have been considered. Three fraction treatments were simulated with displacements in X, Y and Z directions (3F
) and in ±X and Z directions (3F
). For five fractions, the displacements were ±X, ±Y and Z (5F
) resulting in the largest possible spread of the dose distribution. Plans were evaluated by considering the volumes of the reference dose and the 50% isodose, as well as the Paddick conformity index (PCI), the gradient index (GI) and coverage. Based on the change in coverage a margin was evaluated that would have prevented underdosage of the target.
Two fractions displaced up to ±2 mm in opposing directions almost cancel each other out. Minimum dose was reduced by 10% (3F
), 4% (3F
) and 1% (5F
). The main contribution to the underdosage was due to the small systematic error (shift) included in the simulated displacements.The margin required to keep the coverage of the target comparable with the original plan would be 0.5 mm for the 3F
plan, 0.2 mm for the 3F
plan and 0.0 mm for the 5F
for a set-up error of 2 mm.
For fractionated GK treatment of large brain metastases prescribing to the 50% isodose minimizes the effect of random error.For conventional single fraction SRS, the effect of positional uncertainty is always systematic. However, for fractionated treatments, positional uncertainty can be separated into random and systematic components, where random displacements in opposing directions at least partially cancel each other out and thus reduce the overall systematic effect (i.e. from the residual systemat |
---|---|