Loading…

Scanner and kVp dependence of measured CT numbers in the ACR CT phantom

Quality control testing of CT scanners in our region includes a measurement of CT numbers in the American College of Radiology (ACR) CT phantom using a standardized protocol. CT number values are clinically relevant in determining the composition of various tissues in the body. Accuracy is important...

Full description

Saved in:
Bibliographic Details
Published in:Journal of applied clinical medical physics 2013-11, Vol.14 (6), p.338-349
Main Authors: Cropp, Robert J., Seslija, Petar, Tso, David, Thakur, Yogesh
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:Quality control testing of CT scanners in our region includes a measurement of CT numbers in the American College of Radiology (ACR) CT phantom using a standardized protocol. CT number values are clinically relevant in determining the composition of various tissues in the body. Accuracy is important in the characterization of tumors, assessment of coronary calcium, and identification of urinary stone composition. Effective quality control requires that tolerance ranges of CT number values be defined: a measured value outside the range indicates the need for further investigation and possible recalibration of the scanner. This paper presents the results of CT number measurements on 36 scanners (25 GE, 10 Siemens and 1 Toshiba) at each available kVp. Among the five materials (solid water, air, polyethylene, acrylic, bone‐equivalent) the measured CT numbers exhibit manufacturer and kVp dependence, which should be taken into account when defining tolerances. With this scan protocol, air and solid water values are significantly higher on GE scanners than on Siemens scanners (p‐value
ISSN:1526-9914
1526-9914
DOI:10.1120/jacmp.v14i6.4417