Loading…
Evaluation of the efficacy of a medium cut-off dialyser and comparison with other high-flux dialysers in conventional haemodialysis and online haemodiafiltration
ABSTRACT Background Online haemodiafiltration (OL-HDF) has been shown to reduce all-cause mortality versus conventional haemodialysis (HD); however, it is not always available. In these situations, a novel class of membranes with a higher pore size, medium cut-off (MCO) dialysers, could be promising...
Saved in:
Published in: | Clinical kidney journal 2018-10, Vol.11 (5), p.742-746 |
---|---|
Main Authors: | , , , , , , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | ABSTRACT
Background
Online haemodiafiltration (OL-HDF) has been shown to reduce all-cause mortality versus conventional haemodialysis (HD); however, it is not always available. In these situations, a novel class of membranes with a higher pore size, medium cut-off (MCO) dialysers, could be promising. The aim of this study is to evaluate the efficacy of an MCO dialyser in the removal of small and medium-size molecules and compare it with standard high-flux (HF) dialysers in HD and OL-HDF.
Methods
In this crossover study, 18 prevalent HD patients were studied in three single mid-week dialysis treatments during three consecutive weeks as follows: first week with OL-HDF with a standard HF dialyser, second week with conventional HD with a standard HF dialyser and third week with conventional HD with an MCO dialyser. Reduction ratios (RRs) of different-sized molecules and albumin losses were collected for the different dialysers.
Results
MCO HD provided a greater reduction of middle and larger middle molecules compared with standard HF HD [rate reduction (RR) β2-microglobulin 74.7% versus 69.7%, P=0.01; RR myoglobin 62.5% versus 34.3%, P=0.001; RR prolactin 60% versus 32.8%, P=0.001; RR α1-glycoprotein 2.8% versus −0.1%, P=0.01]. We found no difference in the clearance of small and larger middle molecules comparing MCO HD with OL-HDF. Albumin losses were 0.03 g/session with MCO HD and 3.1 g/session with OL-HDF (P=0.001).
Conclusion
MCO HD is superior to standard HF HD in the removal of middle and larger middle molecules and it is not inferior to OL-HDF in the clearance of small and larger middle molecules. Thus it could be an alternative in patients in which it is not possible to perform OL-HDF. |
---|---|
ISSN: | 2048-8505 2048-8513 |
DOI: | 10.1093/ckj/sfy004 |