Development and Validation of a Smartphone-Based Visual Acuity Test (Vision at Home)
To describe the development and validation of a smartphone-based visual acuity (VA) test called Vision at home (V@home). Three study populations (elderly Chinese, adolescent Chinese, and Australian groups) underwent distance and near VA testing using standard Early Treatment Diabetic Retinopathy Stu...
Saved in:
Published in: | Translational vision science & technology 2019-07, Vol.8 (4), p.27-27 |
---|---|
Main Authors: | , , , , , , , , |
Format: | Article |
Language: | English |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | To describe the development and validation of a smartphone-based visual acuity (VA) test called Vision at home (V@home).
Three study populations (elderly Chinese, adolescent Chinese, and Australian groups) underwent distance and near VA testing using standard Early Treatment Diabetic Retinopathy Study (ETDRS) charts and the V@home device; all VA tests used tumbling E optotypes. VA tests were repeated with one eye, selected randomly. Distance VA was measured monocularly at 2 m, and near VA was measured binocularly at 40 cm. Participants also completed a questionnaire about their satisfaction with the device. V@home VA (logMAR) was compared to VA for ETDRS charts at distance and near and test-retest reliability.
The mean difference between V@home and ETDRS distance VA across all groups ranged from -0.010 to -0.100 logMAR. Tolerant weighted kappa (TWK) agreement ranged from substantial (0.742) in the Australian group to almost perfect (0.950) in the adolescent Chinese group. There was high agreement of V@home with near ETDRS VA across all groups, with a mean difference of -0.092 to -0.042 logMAR and a TWK of 0.736 to 0.837. Test-retest reliability was also high (difference: -0.018 to 0.026) for both distance and near VA tests (95% limits of agreement: -0.289 to 0.258 for distance and -0.235 to 0.199 for near). The majority of participants were satisfied with V@home.
V@home could accurately and reliably measure both distance and near VA and is well accepted by participants.
The V@home system could potentially serve as a useful tool to improve eye care accessibility, especially in underdeveloped areas with limited eye care personnel and resources. |
---|---|
ISSN: | 2164-2591 2164-2591 |
DOI: | 10.1167/tvst.8.4.27 |