POWGEN: rebuild of a third‐generation powder diffractometer at the Spallation Neutron Source
The neutron powder diffractometer POWGEN at the Spallation Neutron Source has recently (2017–2018) undergone an upgrade which resulted in an increased detector complement along with a full overhaul of the structural design of the instrument. The current instrument has a solid angular coverage of 1.2...
Saved in:
Published in: | Journal of applied crystallography 2019-10, Vol.52 (5), p.1189-1201 |
---|---|
Main Authors: | , , , , , , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | The neutron powder diffractometer POWGEN at the Spallation Neutron Source has recently (2017–2018) undergone an upgrade which resulted in an increased detector complement along with a full overhaul of the structural design of the instrument. The current instrument has a solid angular coverage of 1.2 steradians and maintains the original third‐generation concept, providing a single‐histogram data set over a wide d‐spacing range and high resolution to access large unit cells, detailed structural refinements and in situ/operando measurements.
This work describes the design principles and upgrade of the neutron powder diffractometer POWGEN at the Spallation Neutron Source. |
---|---|