Loading…
Prospective Evaluation of the mariPOC Test for Detection of Clostridioides difficile Glutamate Dehydrogenase and Toxins A/B
The objective of this study was to evaluate a novel automated random-access test, mariPOC CDI (ArcDia Ltd., Finland), for the detection of glutamate dehydrogenase (GDH) and toxins A and B directly from fecal specimens. The mariPOC test was compared with both the GenomEra PCR assay (Abacus Diagnostic...
Saved in:
Published in: | Journal of clinical microbiology 2020-03, Vol.58 (4) |
---|---|
Main Authors: | , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | The objective of this study was to evaluate a novel automated random-access test, mariPOC CDI (ArcDia Ltd., Finland), for the detection of
glutamate dehydrogenase (GDH) and toxins A and B directly from fecal specimens. The mariPOC test was compared with both the GenomEra
PCR assay (Abacus Diagnostica Oy, Finland) and the TechLab C. diff Quik Chek Complete (Alere Inc.; now Abbot) membrane enzyme immunoassay (MEIA). Culture and the Xpert
assay (Cepheid Inc., USA) were used to resolve discrepant results. In total, 337 specimens were tested with the mariPOC CDI test and GenomEra PCR. Of these specimens, 157 were also tested with the TechLab MEIA. The sensitivity of the mariPOC test for GDH was slightly lower (95.2%) than that obtained with the TechLab assay (100.0%), but no toxin-positive cases were missed. The sensitivity of the mariPOC test for the detection of toxigenic
by analyzing toxin expression was better (81.6%) than that of the TechLab assay (71.1%). The analytical specificities for the mariPOC and the TechLab tests were 98.3% and 100.0% for GDH and 100.0% and 99.2% for toxin A/B, respectively. The analytical specificity of the GenomEra method was 100.0%. The mariPOC and TechLab GDH tests and GenomEra PCR had high negative predictive values of 99.3%, 98.3%, and 99.7%, respectively, in excluding infection with toxigenic
The mariPOC toxin A/B test and GenomEra PCR had an identical analytical positive predictive value of 100%, providing highly reliable information about toxin expression and the presence of toxin genes, respectively. |
---|---|
ISSN: | 0095-1137 1098-660X |
DOI: | 10.1128/JCM.01872-19 |