Assessment and correction of macroscopic field variations in 2D spoiled gradient‐echo sequences
Purpose To model and correct the dephasing effects in the gradient‐echo signal for arbitrary RF excitation pulses with large flip angles in the presence of macroscopic field variations. Methods The dephasing of the spoiled 2D gradient‐echo signal was modeled using a numerical solution of the Bloch e...
Saved in:
Published in: | Magnetic resonance in medicine 2020-08, Vol.84 (2), p.620-633 |
---|---|
Main Authors: | , , , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Purpose
To model and correct the dephasing effects in the gradient‐echo signal for arbitrary RF excitation pulses with large flip angles in the presence of macroscopic field variations.
Methods
The dephasing of the spoiled 2D gradient‐echo signal was modeled using a numerical solution of the Bloch equations to calculate the magnitude and phase of the transverse magnetization across the slice profile. Additionally, regional variations of the transmit RF field and slice profile scaling due to macroscopic field gradients were included. Simulations, phantom, and in vivo measurements at 3 T were conducted for R2∗ and myelin water fraction (MWF) mapping.
Results
The influence of macroscopic field gradients on R2∗ and myelin water fraction estimation can be substantially reduced by applying the proposed model. Moreover, it was shown that the dephasing over time for flip angles of 60° or greater also depends on the polarity of the slice‐selection gradient because of phase variation along the slice profile.
Conclusion
Substantial improvements in R2∗ accuracy and myelin water fraction mapping coverage can be achieved using the proposed model if higher flip angles are required. In this context, we demonstrated that the phase along the slice profile and the polarity of the slice‐selection gradient are essential for proper modeling of the gradient‐echo signal in the presence of macroscopic field variations. |
---|---|