Microdosimetric quantities of an accelerator-based neutron source used for boron neutron capture therapy measured using a gas-filled proportional counter
Abstract Boron neutron capture therapy (BNCT) is an emerging radiation treatment modality, exhibiting the potential to selectively destroy cancer cells. Currently, BNCT is conducted using a nuclear reactor. However, the future trend is to move toward an accelerator-based system for use in hospital e...
Saved in:
Published in: | Journal of radiation research 2020-03, Vol.61 (2), p.214-220 |
---|---|
Main Authors: | , , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Abstract
Boron neutron capture therapy (BNCT) is an emerging radiation treatment modality, exhibiting the potential to selectively destroy cancer cells. Currently, BNCT is conducted using a nuclear reactor. However, the future trend is to move toward an accelerator-based system for use in hospital environments. A typical BNCT radiation field has several different types of radiation. The beam quality should be quantified to accurately determine the dose to be delivered to the target. This study utilized a tissue equivalent proportional counter (TEPC) to measure microdosimetric and macrodosimetric quantities of an accelerator-based neutron source. The micro- and macro-dosimetric quantities measured with the TEPC were compared with those obtained via the the particle and heavy ion transport code system (PHITS) Monte Carlo simulation. The absorbed dose from events >20 keV/μm measured free in air for a 1-h irradiation was calculated as 1.31 ± 0.02 Gy. The simulated result was 1.41 ± 0.07 Gy. The measured and calculated values exhibit good agreement. The relative biological effectiveness (RBE) that was evaluated from the measured microdosimetric spectrum was calculated as 3.7 ± 0.02, similar to the simulated value of 3.8 ± 0.1. These results showed the PHITS Monte Carlo simulation can simulate both micro- and macro-dosimetric quantities accurately. The RBE was calculated using a single-response function, and the results were compared with those of several other institutes that used a similar method. However, care must be taken when using such a single-response function for clinical application, as it is only valid for low doses. For clinical dose ranges (i.e., high doses), multievent distribution inside the target needs to be considered. |
---|---|
ISSN: | 0449-3060 1349-9157 |
DOI: | 10.1093/jrr/rrz101 |