Accuracy of arrhythmia detection in implantable cardiac monitors: A prospective randomized clinical trial comparing Reveal LINQ and Confirm Rx
Background Implantable cardiac monitors (ICMs) are increasingly used to detect arrhythmias in various clinical situations. However, the data transmission time and accuracy of detecting cardiac arrhythmias are unclear. Objective The objective of this study was to compare the efficiency of data transm...
Saved in:
Published in: | Pacing and clinical electrophysiology 2020-11, Vol.43 (11), p.1344-1350 |
---|---|
Main Authors: | , , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Background
Implantable cardiac monitors (ICMs) are increasingly used to detect arrhythmias in various clinical situations. However, the data transmission time and accuracy of detecting cardiac arrhythmias are unclear.
Objective
The objective of this study was to compare the efficiency of data transmission and arrhythmia detection accuracy of the Reveal LINQ with TruRhythm Detection with the Confirm Rx with SharpSense Technology.
Methods
In this prospective study, 142 patients were randomized 1:1 to receive Reveal LINQ or Confirm Rx ICM system. Arrhythmic events include atrial fibrillation (AF), pauses, and bradycardia. Data transmission time is defined as the time from event occurrence to physician notification. All the arrhythmic events are adjudicated for accuracy.
Results
A total of 3510 events were transmitted in 61 patients over 7.1 ± 3.5 months. The transmission time both for all events (448 ± 271 vs 610 ± 515 minutes, P = .02) and for patient activated triggers (24 ± 103 vs 475 ± 426 minutes, P |
---|---|
ISSN: | 0147-8389 1540-8159 |
DOI: | 10.1111/pace.14076 |