Surgical Drill Guide for Insertion of an Infra-Acetabular Screw Based on an Anatomically Precontoured Plate System: A Cadaveric Study

Purpose. Due to the anatomic structure of the pelvis, free-hand placement of screws in the acetabular fracture management can be difficult. Infra-acetabular screw fixation increases acetabular stability by distal fixation of the cup. Aim of this cadaveric study is to investigate if a plate-reference...

Full description

Saved in:
Bibliographic Details
Published in:BioMed research international 2021, Vol.2021, p.1-7
Main Authors: Freigang, Viola, Gottsauner, Maximilian, Rupp, Markus, Pfeifer, Christian, Grechenig, Stephan, Kerner, Alexander, Alt, Volker, Baumann, Florian
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
cited_by cdi_FETCH-LOGICAL-c481t-2ca4e373c14c005f6680959b66c0cedd911f15a9c25c0fe25488f94afe8144df3
cites cdi_FETCH-LOGICAL-c481t-2ca4e373c14c005f6680959b66c0cedd911f15a9c25c0fe25488f94afe8144df3
container_end_page 7
container_issue
container_start_page 1
container_title BioMed research international
container_volume 2021
creator Freigang, Viola
Gottsauner, Maximilian
Rupp, Markus
Pfeifer, Christian
Grechenig, Stephan
Kerner, Alexander
Alt, Volker
Baumann, Florian
description Purpose. Due to the anatomic structure of the pelvis, free-hand placement of screws in the acetabular fracture management can be difficult. Infra-acetabular screw fixation increases acetabular stability by distal fixation of the cup. Aim of this cadaveric study is to investigate if a plate-referenced drill guide can provide save placement of an infra-acetabular screw over a precontoured suprapectineal quadrilateral buttress plate (SQBP). Methods. We constructed a drill guide for an infra-acetabular screw based on the surface of an anatomically precontoured SQBP. A total of 12 adult cadaveric acetabular specimens were used for drill guide-assisted placement of the infra-acetabular screw. The drill guide contains a radiopaque spiral to allow longitudinal fine adjustment of the SQBP along the pelvic brim to assure correct position of the plate-drill-guide construct in relation to the Koehler’s teardrop. After screw placement, we conducted a computed tomography (CT) scan of all specimens to assess the actual position of the screw in relation of the infra-acetabular corridor and the acetabular joint surface. Results. The position of the screw was within the infra-acetabular corridor in all cases. We did not see any intra-articular or intrapelvic screw penetration. The mean distance of the centerline of the screw to the medial border of the infra-acetabular corridor was 3.35 mm. The secure distance to the virtual surface of the femoral head to was 7.3 mm. Conclusions. A plate-referenced drill guide can provide safe placement of an infra-acetabular screw for treatment of acetabular fractures. Radiographic fine adjustment is necessary to access the optimal entry point.
doi_str_mv 10.1155/2021/2321504
format article
fullrecord <record><control><sourceid>gale_pubme</sourceid><recordid>TN_cdi_pubmedcentral_primary_oai_pubmedcentral_nih_gov_8331300</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><galeid>A696876546</galeid><sourcerecordid>A696876546</sourcerecordid><originalsourceid>FETCH-LOGICAL-c481t-2ca4e373c14c005f6680959b66c0cedd911f15a9c25c0fe25488f94afe8144df3</originalsourceid><addsrcrecordid>eNp9kt9qFDEUhwdRbKm98wEC3gg6Nv93xgthXLUWChZWr0M2c7JNySQ1ybTsA_jeZt2lohfNTRLOly854dc0Lwl-R4gQZxRTckYZJQLzJ80xZYS3knDy9GHN2FFzmvMNrqMjEvfyeXPEOBP1BD5ufq3mtHFGe_QpOe_R-exGQDYmdBEypOJiQNEiHereJt0OBopez14ntDIJ7tFHnWFElarIEHSJ087mt-gqgYmhxDnV-pXXBdBqmwtM79GAlnrUd5CcQasyj9sXzTOrfYbTw3zS_Pjy-fvya3v57fxiOVy2hnektNRoDmzBDOEGY2Gl7HAv-rWUBhsYx54QS4TuDRUGW6CCd53tubbQEc5Hy06aD3vv7byeYDQQStJe3SY36bRVUTv1byW4a7WJd6pjjDCMq-D1QZDizxlyUZPLBrzXAeKcFRWi55TLBa3oq__Qm_oXobb3h2Ksvhb_pTbag3LBxnqv2UnVIHvZLaTg8nGqY2LBKO8q9XZPmRRzTmAfGiNY7eKidnFRh7hU_M0ev3Zh1Pfucfo3qcW7PA</addsrcrecordid><sourcetype>Open Access Repository</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>2559339110</pqid></control><display><type>article</type><title>Surgical Drill Guide for Insertion of an Infra-Acetabular Screw Based on an Anatomically Precontoured Plate System: A Cadaveric Study</title><source>Publicly Available Content Database (Proquest) (PQ_SDU_P3)</source><source>Wiley Online Library Open Access</source><creator>Freigang, Viola ; Gottsauner, Maximilian ; Rupp, Markus ; Pfeifer, Christian ; Grechenig, Stephan ; Kerner, Alexander ; Alt, Volker ; Baumann, Florian</creator><contributor>Fortina, Mattia ; Mattia Fortina</contributor><creatorcontrib>Freigang, Viola ; Gottsauner, Maximilian ; Rupp, Markus ; Pfeifer, Christian ; Grechenig, Stephan ; Kerner, Alexander ; Alt, Volker ; Baumann, Florian ; Fortina, Mattia ; Mattia Fortina</creatorcontrib><description>Purpose. Due to the anatomic structure of the pelvis, free-hand placement of screws in the acetabular fracture management can be difficult. Infra-acetabular screw fixation increases acetabular stability by distal fixation of the cup. Aim of this cadaveric study is to investigate if a plate-referenced drill guide can provide save placement of an infra-acetabular screw over a precontoured suprapectineal quadrilateral buttress plate (SQBP). Methods. We constructed a drill guide for an infra-acetabular screw based on the surface of an anatomically precontoured SQBP. A total of 12 adult cadaveric acetabular specimens were used for drill guide-assisted placement of the infra-acetabular screw. The drill guide contains a radiopaque spiral to allow longitudinal fine adjustment of the SQBP along the pelvic brim to assure correct position of the plate-drill-guide construct in relation to the Koehler’s teardrop. After screw placement, we conducted a computed tomography (CT) scan of all specimens to assess the actual position of the screw in relation of the infra-acetabular corridor and the acetabular joint surface. Results. The position of the screw was within the infra-acetabular corridor in all cases. We did not see any intra-articular or intrapelvic screw penetration. The mean distance of the centerline of the screw to the medial border of the infra-acetabular corridor was 3.35 mm. The secure distance to the virtual surface of the femoral head to was 7.3 mm. Conclusions. A plate-referenced drill guide can provide safe placement of an infra-acetabular screw for treatment of acetabular fractures. Radiographic fine adjustment is necessary to access the optimal entry point.</description><identifier>ISSN: 2314-6133</identifier><identifier>EISSN: 2314-6141</identifier><identifier>DOI: 10.1155/2021/2321504</identifier><identifier>PMID: 34355040</identifier><language>eng</language><publisher>New York: Hindawi</publisher><subject>Acetabulum ; Analysis ; Bone implants ; Buttresses ; Cadavers ; Cartilage ; Computed tomography ; Fixation ; Fracture fixation ; Fractures ; Hip joint ; Internal fixation in fractures ; Management ; Methods ; Pelvis ; Quadrilaterals ; Software ; Statistical analysis ; Surgical techniques</subject><ispartof>BioMed research international, 2021, Vol.2021, p.1-7</ispartof><rights>Copyright © 2021 Viola Freigang et al.</rights><rights>COPYRIGHT 2021 John Wiley &amp; Sons, Inc.</rights><rights>Copyright © 2021 Viola Freigang et al. This is an open access article distributed under the Creative Commons Attribution License (the “License”), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License. https://creativecommons.org/licenses/by/4.0</rights><rights>Copyright © 2021 Viola Freigang et al. 2021</rights><lds50>peer_reviewed</lds50><oa>free_for_read</oa><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c481t-2ca4e373c14c005f6680959b66c0cedd911f15a9c25c0fe25488f94afe8144df3</citedby><cites>FETCH-LOGICAL-c481t-2ca4e373c14c005f6680959b66c0cedd911f15a9c25c0fe25488f94afe8144df3</cites><orcidid>0000-0003-1413-3587</orcidid></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><linktopdf>$$Uhttps://www.proquest.com/docview/2559339110/fulltextPDF?pq-origsite=primo$$EPDF$$P50$$Gproquest$$Hfree_for_read</linktopdf><linktohtml>$$Uhttps://www.proquest.com/docview/2559339110?pq-origsite=primo$$EHTML$$P50$$Gproquest$$Hfree_for_read</linktohtml><link.rule.ids>230,314,780,784,885,4024,25753,27923,27924,27925,37012,37013,44590,75126</link.rule.ids></links><search><contributor>Fortina, Mattia</contributor><contributor>Mattia Fortina</contributor><creatorcontrib>Freigang, Viola</creatorcontrib><creatorcontrib>Gottsauner, Maximilian</creatorcontrib><creatorcontrib>Rupp, Markus</creatorcontrib><creatorcontrib>Pfeifer, Christian</creatorcontrib><creatorcontrib>Grechenig, Stephan</creatorcontrib><creatorcontrib>Kerner, Alexander</creatorcontrib><creatorcontrib>Alt, Volker</creatorcontrib><creatorcontrib>Baumann, Florian</creatorcontrib><title>Surgical Drill Guide for Insertion of an Infra-Acetabular Screw Based on an Anatomically Precontoured Plate System: A Cadaveric Study</title><title>BioMed research international</title><description>Purpose. Due to the anatomic structure of the pelvis, free-hand placement of screws in the acetabular fracture management can be difficult. Infra-acetabular screw fixation increases acetabular stability by distal fixation of the cup. Aim of this cadaveric study is to investigate if a plate-referenced drill guide can provide save placement of an infra-acetabular screw over a precontoured suprapectineal quadrilateral buttress plate (SQBP). Methods. We constructed a drill guide for an infra-acetabular screw based on the surface of an anatomically precontoured SQBP. A total of 12 adult cadaveric acetabular specimens were used for drill guide-assisted placement of the infra-acetabular screw. The drill guide contains a radiopaque spiral to allow longitudinal fine adjustment of the SQBP along the pelvic brim to assure correct position of the plate-drill-guide construct in relation to the Koehler’s teardrop. After screw placement, we conducted a computed tomography (CT) scan of all specimens to assess the actual position of the screw in relation of the infra-acetabular corridor and the acetabular joint surface. Results. The position of the screw was within the infra-acetabular corridor in all cases. We did not see any intra-articular or intrapelvic screw penetration. The mean distance of the centerline of the screw to the medial border of the infra-acetabular corridor was 3.35 mm. The secure distance to the virtual surface of the femoral head to was 7.3 mm. Conclusions. A plate-referenced drill guide can provide safe placement of an infra-acetabular screw for treatment of acetabular fractures. Radiographic fine adjustment is necessary to access the optimal entry point.</description><subject>Acetabulum</subject><subject>Analysis</subject><subject>Bone implants</subject><subject>Buttresses</subject><subject>Cadavers</subject><subject>Cartilage</subject><subject>Computed tomography</subject><subject>Fixation</subject><subject>Fracture fixation</subject><subject>Fractures</subject><subject>Hip joint</subject><subject>Internal fixation in fractures</subject><subject>Management</subject><subject>Methods</subject><subject>Pelvis</subject><subject>Quadrilaterals</subject><subject>Software</subject><subject>Statistical analysis</subject><subject>Surgical techniques</subject><issn>2314-6133</issn><issn>2314-6141</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2021</creationdate><recordtype>article</recordtype><sourceid>PIMPY</sourceid><recordid>eNp9kt9qFDEUhwdRbKm98wEC3gg6Nv93xgthXLUWChZWr0M2c7JNySQ1ybTsA_jeZt2lohfNTRLOly854dc0Lwl-R4gQZxRTckYZJQLzJ80xZYS3knDy9GHN2FFzmvMNrqMjEvfyeXPEOBP1BD5ufq3mtHFGe_QpOe_R-exGQDYmdBEypOJiQNEiHereJt0OBopez14ntDIJ7tFHnWFElarIEHSJ087mt-gqgYmhxDnV-pXXBdBqmwtM79GAlnrUd5CcQasyj9sXzTOrfYbTw3zS_Pjy-fvya3v57fxiOVy2hnektNRoDmzBDOEGY2Gl7HAv-rWUBhsYx54QS4TuDRUGW6CCd53tubbQEc5Hy06aD3vv7byeYDQQStJe3SY36bRVUTv1byW4a7WJd6pjjDCMq-D1QZDizxlyUZPLBrzXAeKcFRWi55TLBa3oq__Qm_oXobb3h2Ksvhb_pTbag3LBxnqv2UnVIHvZLaTg8nGqY2LBKO8q9XZPmRRzTmAfGiNY7eKidnFRh7hU_M0ev3Zh1Pfucfo3qcW7PA</recordid><startdate>2021</startdate><enddate>2021</enddate><creator>Freigang, Viola</creator><creator>Gottsauner, Maximilian</creator><creator>Rupp, Markus</creator><creator>Pfeifer, Christian</creator><creator>Grechenig, Stephan</creator><creator>Kerner, Alexander</creator><creator>Alt, Volker</creator><creator>Baumann, Florian</creator><general>Hindawi</general><general>John Wiley &amp; Sons, Inc</general><general>Hindawi Limited</general><scope>RHU</scope><scope>RHW</scope><scope>RHX</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>3V.</scope><scope>7QL</scope><scope>7QO</scope><scope>7T7</scope><scope>7TK</scope><scope>7U7</scope><scope>7U9</scope><scope>7X7</scope><scope>7XB</scope><scope>88E</scope><scope>8FD</scope><scope>8FE</scope><scope>8FG</scope><scope>8FH</scope><scope>8FI</scope><scope>8FJ</scope><scope>8FK</scope><scope>ABUWG</scope><scope>AFKRA</scope><scope>ARAPS</scope><scope>AZQEC</scope><scope>BBNVY</scope><scope>BENPR</scope><scope>BGLVJ</scope><scope>BHPHI</scope><scope>C1K</scope><scope>CCPQU</scope><scope>CWDGH</scope><scope>DWQXO</scope><scope>FR3</scope><scope>FYUFA</scope><scope>GHDGH</scope><scope>GNUQQ</scope><scope>H94</scope><scope>HCIFZ</scope><scope>K9.</scope><scope>LK8</scope><scope>M0S</scope><scope>M1P</scope><scope>M7N</scope><scope>M7P</scope><scope>P5Z</scope><scope>P62</scope><scope>P64</scope><scope>PIMPY</scope><scope>PQEST</scope><scope>PQQKQ</scope><scope>PQUKI</scope><scope>7X8</scope><scope>5PM</scope><orcidid>https://orcid.org/0000-0003-1413-3587</orcidid></search><sort><creationdate>2021</creationdate><title>Surgical Drill Guide for Insertion of an Infra-Acetabular Screw Based on an Anatomically Precontoured Plate System: A Cadaveric Study</title><author>Freigang, Viola ; Gottsauner, Maximilian ; Rupp, Markus ; Pfeifer, Christian ; Grechenig, Stephan ; Kerner, Alexander ; Alt, Volker ; Baumann, Florian</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c481t-2ca4e373c14c005f6680959b66c0cedd911f15a9c25c0fe25488f94afe8144df3</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2021</creationdate><topic>Acetabulum</topic><topic>Analysis</topic><topic>Bone implants</topic><topic>Buttresses</topic><topic>Cadavers</topic><topic>Cartilage</topic><topic>Computed tomography</topic><topic>Fixation</topic><topic>Fracture fixation</topic><topic>Fractures</topic><topic>Hip joint</topic><topic>Internal fixation in fractures</topic><topic>Management</topic><topic>Methods</topic><topic>Pelvis</topic><topic>Quadrilaterals</topic><topic>Software</topic><topic>Statistical analysis</topic><topic>Surgical techniques</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Freigang, Viola</creatorcontrib><creatorcontrib>Gottsauner, Maximilian</creatorcontrib><creatorcontrib>Rupp, Markus</creatorcontrib><creatorcontrib>Pfeifer, Christian</creatorcontrib><creatorcontrib>Grechenig, Stephan</creatorcontrib><creatorcontrib>Kerner, Alexander</creatorcontrib><creatorcontrib>Alt, Volker</creatorcontrib><creatorcontrib>Baumann, Florian</creatorcontrib><collection>Hindawi Publishing Complete</collection><collection>Hindawi Publishing Subscription Journals</collection><collection>Hindawi Publishing Open Access</collection><collection>CrossRef</collection><collection>ProQuest Central (Corporate)</collection><collection>Bacteriology Abstracts (Microbiology B)</collection><collection>Biotechnology Research Abstracts</collection><collection>Industrial and Applied Microbiology Abstracts (Microbiology A)</collection><collection>Neurosciences Abstracts</collection><collection>Toxicology Abstracts</collection><collection>Virology and AIDS Abstracts</collection><collection>ProQuest Health and Medical</collection><collection>ProQuest Central (purchase pre-March 2016)</collection><collection>Medical Database (Alumni Edition)</collection><collection>Technology Research Database</collection><collection>ProQuest SciTech Collection</collection><collection>ProQuest Technology Collection</collection><collection>ProQuest Natural Science Collection</collection><collection>Hospital Premium Collection</collection><collection>Hospital Premium Collection (Alumni Edition)</collection><collection>ProQuest Central (Alumni) (purchase pre-March 2016)</collection><collection>ProQuest Central (Alumni)</collection><collection>ProQuest Central</collection><collection>Advanced Technologies &amp; Aerospace Collection</collection><collection>ProQuest Central Essentials</collection><collection>Biological Science Collection</collection><collection>AUTh Library subscriptions: ProQuest Central</collection><collection>Technology Collection</collection><collection>ProQuest Natural Science Collection</collection><collection>Environmental Sciences and Pollution Management</collection><collection>ProQuest One Community College</collection><collection>Middle East &amp; Africa Database</collection><collection>ProQuest Central</collection><collection>Engineering Research Database</collection><collection>Health Research Premium Collection</collection><collection>Health Research Premium Collection (Alumni)</collection><collection>ProQuest Central Student</collection><collection>AIDS and Cancer Research Abstracts</collection><collection>SciTech Premium Collection (Proquest) (PQ_SDU_P3)</collection><collection>ProQuest Health &amp; Medical Complete (Alumni)</collection><collection>ProQuest Biological Science Collection</collection><collection>Health &amp; Medical Collection (Alumni Edition)</collection><collection>PML(ProQuest Medical Library)</collection><collection>Algology Mycology and Protozoology Abstracts (Microbiology C)</collection><collection>ProQuest Biological Science Journals</collection><collection>ProQuest advanced technologies &amp; aerospace journals</collection><collection>ProQuest Advanced Technologies &amp; Aerospace Collection</collection><collection>Biotechnology and BioEngineering Abstracts</collection><collection>Publicly Available Content Database (Proquest) (PQ_SDU_P3)</collection><collection>ProQuest One Academic Eastern Edition (DO NOT USE)</collection><collection>ProQuest One Academic</collection><collection>ProQuest One Academic UKI Edition</collection><collection>MEDLINE - Academic</collection><collection>PubMed Central (Full Participant titles)</collection><jtitle>BioMed research international</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Freigang, Viola</au><au>Gottsauner, Maximilian</au><au>Rupp, Markus</au><au>Pfeifer, Christian</au><au>Grechenig, Stephan</au><au>Kerner, Alexander</au><au>Alt, Volker</au><au>Baumann, Florian</au><au>Fortina, Mattia</au><au>Mattia Fortina</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Surgical Drill Guide for Insertion of an Infra-Acetabular Screw Based on an Anatomically Precontoured Plate System: A Cadaveric Study</atitle><jtitle>BioMed research international</jtitle><date>2021</date><risdate>2021</risdate><volume>2021</volume><spage>1</spage><epage>7</epage><pages>1-7</pages><issn>2314-6133</issn><eissn>2314-6141</eissn><abstract>Purpose. Due to the anatomic structure of the pelvis, free-hand placement of screws in the acetabular fracture management can be difficult. Infra-acetabular screw fixation increases acetabular stability by distal fixation of the cup. Aim of this cadaveric study is to investigate if a plate-referenced drill guide can provide save placement of an infra-acetabular screw over a precontoured suprapectineal quadrilateral buttress plate (SQBP). Methods. We constructed a drill guide for an infra-acetabular screw based on the surface of an anatomically precontoured SQBP. A total of 12 adult cadaveric acetabular specimens were used for drill guide-assisted placement of the infra-acetabular screw. The drill guide contains a radiopaque spiral to allow longitudinal fine adjustment of the SQBP along the pelvic brim to assure correct position of the plate-drill-guide construct in relation to the Koehler’s teardrop. After screw placement, we conducted a computed tomography (CT) scan of all specimens to assess the actual position of the screw in relation of the infra-acetabular corridor and the acetabular joint surface. Results. The position of the screw was within the infra-acetabular corridor in all cases. We did not see any intra-articular or intrapelvic screw penetration. The mean distance of the centerline of the screw to the medial border of the infra-acetabular corridor was 3.35 mm. The secure distance to the virtual surface of the femoral head to was 7.3 mm. Conclusions. A plate-referenced drill guide can provide safe placement of an infra-acetabular screw for treatment of acetabular fractures. Radiographic fine adjustment is necessary to access the optimal entry point.</abstract><cop>New York</cop><pub>Hindawi</pub><pmid>34355040</pmid><doi>10.1155/2021/2321504</doi><tpages>7</tpages><orcidid>https://orcid.org/0000-0003-1413-3587</orcidid><oa>free_for_read</oa></addata></record>
fulltext fulltext
identifier ISSN: 2314-6133
ispartof BioMed research international, 2021, Vol.2021, p.1-7
issn 2314-6133
2314-6141
language eng
recordid cdi_pubmedcentral_primary_oai_pubmedcentral_nih_gov_8331300
source Publicly Available Content Database (Proquest) (PQ_SDU_P3); Wiley Online Library Open Access
subjects Acetabulum
Analysis
Bone implants
Buttresses
Cadavers
Cartilage
Computed tomography
Fixation
Fracture fixation
Fractures
Hip joint
Internal fixation in fractures
Management
Methods
Pelvis
Quadrilaterals
Software
Statistical analysis
Surgical techniques
title Surgical Drill Guide for Insertion of an Infra-Acetabular Screw Based on an Anatomically Precontoured Plate System: A Cadaveric Study
url http://sfxeu10.hosted.exlibrisgroup.com/loughborough?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2024-12-26T14%3A13%3A47IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-gale_pubme&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Surgical%20Drill%20Guide%20for%20Insertion%20of%20an%20Infra-Acetabular%20Screw%20Based%20on%20an%20Anatomically%20Precontoured%20Plate%20System:%20A%20Cadaveric%20Study&rft.jtitle=BioMed%20research%20international&rft.au=Freigang,%20Viola&rft.date=2021&rft.volume=2021&rft.spage=1&rft.epage=7&rft.pages=1-7&rft.issn=2314-6133&rft.eissn=2314-6141&rft_id=info:doi/10.1155/2021/2321504&rft_dat=%3Cgale_pubme%3EA696876546%3C/gale_pubme%3E%3Cgrp_id%3Ecdi_FETCH-LOGICAL-c481t-2ca4e373c14c005f6680959b66c0cedd911f15a9c25c0fe25488f94afe8144df3%3C/grp_id%3E%3Coa%3E%3C/oa%3E%3Curl%3E%3C/url%3E&rft_id=info:oai/&rft_pqid=2559339110&rft_id=info:pmid/34355040&rft_galeid=A696876546&rfr_iscdi=true