Evaluation of the Roche-SD Biosensor rapid antigen test: Antigen is not reliable in detecting SARS-CoV-2 at the early stage of infection with respiratory symptoms
We evaluated a rapid antigen test against SARS-CoV-2 virus (Roche-SD Biosensor; RSDB-RAT) in children and adults with respiratory symptoms compared to those with nonrespiratory symptoms or asymptomatic. Also the performance of RSDB-RAT with respect to the duration of respiratory symptoms was assesse...
Saved in:
Published in: | Diagnostic microbiology and infectious disease 2022-04, Vol.102 (4), p.115628-115628, Article 115628 |
---|---|
Main Authors: | , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | We evaluated a rapid antigen test against SARS-CoV-2 virus (Roche-SD Biosensor; RSDB-RAT) in children and adults with respiratory symptoms compared to those with nonrespiratory symptoms or asymptomatic. Also the performance of RSDB-RAT with respect to the duration of respiratory symptoms was assessed. A viral cross-reactivity panel was included. RSDB-RAT was reliable in detecting SARS-CoV-2 in children and adults if the respiratory symptoms had endured 1 to 7 days. If the respiratory symptoms had lasted less than 1 day, the sensitivity was significantly lower. No cross-reactivity with other respiratory viruses was observed. |
---|---|
ISSN: | 0732-8893 1879-0070 |
DOI: | 10.1016/j.diagmicrobio.2021.115628 |