Loading…

Evaluation of a digestion procedure based on the use of diluted nitric acid solutions and H2O2 for the multielement determination of whole milk powder and bovine liver by ICP-based techniquesElectronic supplementary information (ESI) available. See DOI: 10.1039/c3ja50330e

Many efforts have been made in order to reduce the amount of reagents and waste produced in analytical laboratories. However, suitable digestion efficiency must be considered, and depending on the sample preparation procedure, incomplete digestion can result in severe matrix effects during analysis...

Full description

Saved in:
Bibliographic Details
Main Authors: Bizzi, Cezar A, Flores, Eder L. M, Nóbrega, Joaquim A, Oliveira, Jussiane S. S, Schmidt, Lucas, Mortari, Sérgio R
Format: Article
Language:English
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:Many efforts have been made in order to reduce the amount of reagents and waste produced in analytical laboratories. However, suitable digestion efficiency must be considered, and depending on the sample preparation procedure, incomplete digestion can result in severe matrix effects during analysis by spectrometric techniques such as ICP OES and ICP-MS. In the present work a procedure based on the use of H 2 O 2 was developed in order to minimize the consumption of HNO 3 without decreasing the efficiency of digestion. Although H 2 O 2 has been used combined with HNO 3 for sample digestion, its role is still not completely elucidated even as its action combined with O 2 in pressurized systems. The performance obtained using H 2 O 2 was similar to that observed when adding O 2 to the reaction vessel, driving the better understanding of the role of H 2 O 2 in closed digestion procedures. Digestion using H 2 O 2 allowed the use of HNO 3 solutions as diluted as 1 mol L −1 to digest sample masses of biological materials as high as 500 mg. The proposed procedure allowed a reduction of up to 14 and 9.3-fold in the HNO 3 amount normally used in whole milk powder and bovine liver digestions, respectively, without decreasing the digestion efficiency. Calcium, Cu, Fe, K, Mg, Mn, Mo, Na, and Zn were determined by ICP OES, while Cd, Co, and Pb were determined by ICP-MS. Using diluted HNO 3 solution low blank values were obtained resulting in relatively lower limits of detection and relative standard deviations. The accuracy was evaluated by using certified reference materials of milk powder and bovine liver (agreement was better than 95% to certified values for all evaluated analytes). The use of diluted HNO 3 combined with H 2 O 2 for microwave assisted wet digestion allowed for metal determination with minimized matrix effects and low blank values.
ISSN:0267-9477
1364-5544
DOI:10.1039/c3ja50330e