Loading…
Amperometric biosensor precision improvement: application to organophosphorus pesticide determination
An approach for precision improvement in organophosphorus pesticide determination was developed, applied and characterized. The enzyme organophosphorus hydrolase was employed as a specific agent for electrode surface modification, and the pulsed amperometric detection for 66.6 ms was applied for the...
Saved in:
Published in: | Analytical methods 2014-01, Vol.6 (2), p.8232-8238 |
---|---|
Main Authors: | , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | An approach for precision improvement in organophosphorus pesticide determination was developed, applied and characterized. The enzyme organophosphorus hydrolase was employed as a specific agent for electrode surface modification, and the pulsed amperometric detection for 66.6 ms was applied for the hydrolysis product detection. Pulsed electrochemical cleaning,
i.e.
oxidation at a very positive potential (+1.4 V
vs.
Ag, AgCl) was applied every 233 ms for 166.6 ms for elimination of the fouling effect, which degrades the determination precision. By this way a RSD of 0.41% at 40 μmol L
−1
p
-nitrophenol determination was achieved after 10 determinations, instead of 43.93% without the pulsed cleaning application.
An approach for precision improvement in organophosphorus pesticide determination was developed, applied and characterized. |
---|---|
ISSN: | 1759-9660 1759-9679 |
DOI: | 10.1039/c4ay01792g |