Evaluation of tensile strength after insertions andremovals of abutments on frictional Morse taper implants

This study aimed to evaluate possible changes in final retention after nine sequences of insertion and removal (SIR) of a frictional Morse taper implant/abutment system, evaluating the force required for dissociating this set between sequences, and verifying possible deformations in the implant head...

Full description

Saved in:
Bibliographic Details
Published in:International journal of odontostomatology 2021-06, Vol.15 (2), p.356-362
Main Authors: da-Silva, Jader Moreira, Sobral, Brenda Santos, Cabral, Carolina Takis, Souza, Cíntia de Melo Silva, Fonseca, Ricardo Roberto de Souza, Ramacciato, Juliana Cama
Format: Article
Language:English
Subjects:
Citations: Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
cited_by cdi_FETCH-LOGICAL-c1906-a7bced9b5fff264f73e6755dee230e6f69faf2e06b41e03a9f4d84fa89a861a63
cites
container_end_page 362
container_issue 2
container_start_page 356
container_title International journal of odontostomatology
container_volume 15
creator da-Silva, Jader Moreira
Sobral, Brenda Santos
Cabral, Carolina Takis
Souza, Cíntia de Melo Silva
Fonseca, Ricardo Roberto de Souza
Ramacciato, Juliana Cama
description This study aimed to evaluate possible changes in final retention after nine sequences of insertion and removal (SIR) of a frictional Morse taper implant/abutment system, evaluating the force required for dissociating this set between sequences, and verifying possible deformations in the implant heads. Ten implants, 13 mm long and 3.3 mm in diameter, were coupled to a universal mechanical testing machine. Ten anti-rotational abutments, 13 mm long and 3.5 mm in diameter, were connected to the implants parallel to the long axis, using an instrument called beat-connection, and subjected to tensile tests and SEM analysis. The results were analyzed using the Kruskal-Wallis test with Dunn’s post-test, and the significance level was set at 5 %. There was no statistically significant difference in final retention among the nine SIRs evaluated. The force needed to uncouple the abutment from the implant increased as SIRs were performed on all ten implants, and an increase of 29.03 % was observed in the ninth SIR compared to the first SIR. After SEM analysis, no significant deformations, fractures, or cracks were observed in the implant heads.
doi_str_mv 10.4067/S0718-381X2021000200356
format article
fullrecord <record><control><sourceid>scielo_cross</sourceid><recordid>TN_cdi_scielo_journals_S0718_381X2021000200356</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><scielo_id>S0718_381X2021000200356</scielo_id><sourcerecordid>S0718_381X2021000200356</sourcerecordid><originalsourceid>FETCH-LOGICAL-c1906-a7bced9b5fff264f73e6755dee230e6f69faf2e06b41e03a9f4d84fa89a861a63</originalsourceid><addsrcrecordid>eNp1kNtKxDAQhoMouKz7DOYFuk56SNtLWdYDrHihgndl2k402iZLkhV8e9NdEUG8mhnm-4bhZ-xcwDIHWV48QCmqJKvEcwqpAIAUICvkEZv9LI5_9ads4b1uAapU5pGesff1Bw47DNoabhUPZLweiPvgyLyEV44qkOPaeHIT4zma3tFoo-UnAdtdGMmEOBiunO4mCgd-Z50nHnA72eN2wIicsRMVNVp81zl7ulo_rm6Szf317epyk3SiBplg2XbU122hlIpvqjIjWRZFT5RmQFLJWqFKCWSbC4IMa5X3Va6wqrGSAmU2Z8vDXd9pGmzzZncu_uSbfVrNn7SiUB6EzlnvHalm6_SI7rMR0Ew5_2t-AdFhcdk</addsrcrecordid><sourcetype>Open Access Repository</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype></control><display><type>article</type><title>Evaluation of tensile strength after insertions andremovals of abutments on frictional Morse taper implants</title><source>SciELO Chile</source><creator>da-Silva, Jader Moreira ; Sobral, Brenda Santos ; Cabral, Carolina Takis ; Souza, Cíntia de Melo Silva ; Fonseca, Ricardo Roberto de Souza ; Ramacciato, Juliana Cama</creator><creatorcontrib>da-Silva, Jader Moreira ; Sobral, Brenda Santos ; Cabral, Carolina Takis ; Souza, Cíntia de Melo Silva ; Fonseca, Ricardo Roberto de Souza ; Ramacciato, Juliana Cama</creatorcontrib><description>This study aimed to evaluate possible changes in final retention after nine sequences of insertion and removal (SIR) of a frictional Morse taper implant/abutment system, evaluating the force required for dissociating this set between sequences, and verifying possible deformations in the implant heads. Ten implants, 13 mm long and 3.3 mm in diameter, were coupled to a universal mechanical testing machine. Ten anti-rotational abutments, 13 mm long and 3.5 mm in diameter, were connected to the implants parallel to the long axis, using an instrument called beat-connection, and subjected to tensile tests and SEM analysis. The results were analyzed using the Kruskal-Wallis test with Dunn’s post-test, and the significance level was set at 5 %. There was no statistically significant difference in final retention among the nine SIRs evaluated. The force needed to uncouple the abutment from the implant increased as SIRs were performed on all ten implants, and an increase of 29.03 % was observed in the ninth SIR compared to the first SIR. After SEM analysis, no significant deformations, fractures, or cracks were observed in the implant heads.</description><identifier>ISSN: 0718-381X</identifier><identifier>EISSN: 0718-381X</identifier><identifier>DOI: 10.4067/S0718-381X2021000200356</identifier><language>eng</language><publisher>Universidad de La Frontera. Facultad de Medicina</publisher><subject>DENTISTRY, ORAL SURGERY &amp; MEDICINE</subject><ispartof>International journal of odontostomatology, 2021-06, Vol.15 (2), p.356-362</ispartof><rights>This work is licensed under a Creative Commons Attribution-NonCommercial 4.0 International License.</rights><oa>free_for_read</oa><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c1906-a7bced9b5fff264f73e6755dee230e6f69faf2e06b41e03a9f4d84fa89a861a63</citedby></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><link.rule.ids>230,314,780,784,885,24151,27924,27925</link.rule.ids></links><search><creatorcontrib>da-Silva, Jader Moreira</creatorcontrib><creatorcontrib>Sobral, Brenda Santos</creatorcontrib><creatorcontrib>Cabral, Carolina Takis</creatorcontrib><creatorcontrib>Souza, Cíntia de Melo Silva</creatorcontrib><creatorcontrib>Fonseca, Ricardo Roberto de Souza</creatorcontrib><creatorcontrib>Ramacciato, Juliana Cama</creatorcontrib><title>Evaluation of tensile strength after insertions andremovals of abutments on frictional Morse taper implants</title><title>International journal of odontostomatology</title><addtitle>Int. J. Odontostomat</addtitle><description>This study aimed to evaluate possible changes in final retention after nine sequences of insertion and removal (SIR) of a frictional Morse taper implant/abutment system, evaluating the force required for dissociating this set between sequences, and verifying possible deformations in the implant heads. Ten implants, 13 mm long and 3.3 mm in diameter, were coupled to a universal mechanical testing machine. Ten anti-rotational abutments, 13 mm long and 3.5 mm in diameter, were connected to the implants parallel to the long axis, using an instrument called beat-connection, and subjected to tensile tests and SEM analysis. The results were analyzed using the Kruskal-Wallis test with Dunn’s post-test, and the significance level was set at 5 %. There was no statistically significant difference in final retention among the nine SIRs evaluated. The force needed to uncouple the abutment from the implant increased as SIRs were performed on all ten implants, and an increase of 29.03 % was observed in the ninth SIR compared to the first SIR. After SEM analysis, no significant deformations, fractures, or cracks were observed in the implant heads.</description><subject>DENTISTRY, ORAL SURGERY &amp; MEDICINE</subject><issn>0718-381X</issn><issn>0718-381X</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2021</creationdate><recordtype>article</recordtype><recordid>eNp1kNtKxDAQhoMouKz7DOYFuk56SNtLWdYDrHihgndl2k402iZLkhV8e9NdEUG8mhnm-4bhZ-xcwDIHWV48QCmqJKvEcwqpAIAUICvkEZv9LI5_9ads4b1uAapU5pGesff1Bw47DNoabhUPZLweiPvgyLyEV44qkOPaeHIT4zma3tFoo-UnAdtdGMmEOBiunO4mCgd-Z50nHnA72eN2wIicsRMVNVp81zl7ulo_rm6Szf317epyk3SiBplg2XbU122hlIpvqjIjWRZFT5RmQFLJWqFKCWSbC4IMa5X3Va6wqrGSAmU2Z8vDXd9pGmzzZncu_uSbfVrNn7SiUB6EzlnvHalm6_SI7rMR0Ew5_2t-AdFhcdk</recordid><startdate>202106</startdate><enddate>202106</enddate><creator>da-Silva, Jader Moreira</creator><creator>Sobral, Brenda Santos</creator><creator>Cabral, Carolina Takis</creator><creator>Souza, Cíntia de Melo Silva</creator><creator>Fonseca, Ricardo Roberto de Souza</creator><creator>Ramacciato, Juliana Cama</creator><general>Universidad de La Frontera. Facultad de Medicina</general><scope>AAYXX</scope><scope>CITATION</scope><scope>GPN</scope></search><sort><creationdate>202106</creationdate><title>Evaluation of tensile strength after insertions andremovals of abutments on frictional Morse taper implants</title><author>da-Silva, Jader Moreira ; Sobral, Brenda Santos ; Cabral, Carolina Takis ; Souza, Cíntia de Melo Silva ; Fonseca, Ricardo Roberto de Souza ; Ramacciato, Juliana Cama</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c1906-a7bced9b5fff264f73e6755dee230e6f69faf2e06b41e03a9f4d84fa89a861a63</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2021</creationdate><topic>DENTISTRY, ORAL SURGERY &amp; MEDICINE</topic><toplevel>online_resources</toplevel><creatorcontrib>da-Silva, Jader Moreira</creatorcontrib><creatorcontrib>Sobral, Brenda Santos</creatorcontrib><creatorcontrib>Cabral, Carolina Takis</creatorcontrib><creatorcontrib>Souza, Cíntia de Melo Silva</creatorcontrib><creatorcontrib>Fonseca, Ricardo Roberto de Souza</creatorcontrib><creatorcontrib>Ramacciato, Juliana Cama</creatorcontrib><collection>CrossRef</collection><collection>SciELO</collection><jtitle>International journal of odontostomatology</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>da-Silva, Jader Moreira</au><au>Sobral, Brenda Santos</au><au>Cabral, Carolina Takis</au><au>Souza, Cíntia de Melo Silva</au><au>Fonseca, Ricardo Roberto de Souza</au><au>Ramacciato, Juliana Cama</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Evaluation of tensile strength after insertions andremovals of abutments on frictional Morse taper implants</atitle><jtitle>International journal of odontostomatology</jtitle><addtitle>Int. J. Odontostomat</addtitle><date>2021-06</date><risdate>2021</risdate><volume>15</volume><issue>2</issue><spage>356</spage><epage>362</epage><pages>356-362</pages><issn>0718-381X</issn><eissn>0718-381X</eissn><abstract>This study aimed to evaluate possible changes in final retention after nine sequences of insertion and removal (SIR) of a frictional Morse taper implant/abutment system, evaluating the force required for dissociating this set between sequences, and verifying possible deformations in the implant heads. Ten implants, 13 mm long and 3.3 mm in diameter, were coupled to a universal mechanical testing machine. Ten anti-rotational abutments, 13 mm long and 3.5 mm in diameter, were connected to the implants parallel to the long axis, using an instrument called beat-connection, and subjected to tensile tests and SEM analysis. The results were analyzed using the Kruskal-Wallis test with Dunn’s post-test, and the significance level was set at 5 %. There was no statistically significant difference in final retention among the nine SIRs evaluated. The force needed to uncouple the abutment from the implant increased as SIRs were performed on all ten implants, and an increase of 29.03 % was observed in the ninth SIR compared to the first SIR. After SEM analysis, no significant deformations, fractures, or cracks were observed in the implant heads.</abstract><pub>Universidad de La Frontera. Facultad de Medicina</pub><doi>10.4067/S0718-381X2021000200356</doi><tpages>7</tpages><oa>free_for_read</oa></addata></record>
fulltext fulltext
identifier ISSN: 0718-381X
ispartof International journal of odontostomatology, 2021-06, Vol.15 (2), p.356-362
issn 0718-381X
0718-381X
language eng
recordid cdi_scielo_journals_S0718_381X2021000200356
source SciELO Chile
subjects DENTISTRY, ORAL SURGERY & MEDICINE
title Evaluation of tensile strength after insertions andremovals of abutments on frictional Morse taper implants
url http://sfxeu10.hosted.exlibrisgroup.com/loughborough?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2024-12-29T09%3A22%3A48IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-scielo_cross&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Evaluation%20of%20tensile%20strength%20after%20insertions%20andremovals%20of%20abutments%20on%20frictional%20Morse%20taper%20implants&rft.jtitle=International%20journal%20of%20odontostomatology&rft.au=da-Silva,%20Jader%20Moreira&rft.date=2021-06&rft.volume=15&rft.issue=2&rft.spage=356&rft.epage=362&rft.pages=356-362&rft.issn=0718-381X&rft.eissn=0718-381X&rft_id=info:doi/10.4067/S0718-381X2021000200356&rft_dat=%3Cscielo_cross%3ES0718_381X2021000200356%3C/scielo_cross%3E%3Cgrp_id%3Ecdi_FETCH-LOGICAL-c1906-a7bced9b5fff264f73e6755dee230e6f69faf2e06b41e03a9f4d84fa89a861a63%3C/grp_id%3E%3Coa%3E%3C/oa%3E%3Curl%3E%3C/url%3E&rft_id=info:oai/&rft_id=info:pmid/&rft_scielo_id=S0718_381X2021000200356&rfr_iscdi=true