New method for very sensitive dielectric difference measurements on high‐loss liquids at microwave frequencies

A microwave system for very sensitive dielectric measurements on high‐loss liquids over the frequency range 8–26.5 GHz is described which depends on the cavity resonance perturbation method. Equations for the resonance perturbation by a thin capillary in cylindrical as well as in rectangular cavitie...

Full description

Saved in:
Bibliographic Details
Published in:Review of scientific instruments 1975-12, Vol.46 (12), p.1691-1696
Main Author: Hallenga, K.
Format: Article
Language:English
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:A microwave system for very sensitive dielectric measurements on high‐loss liquids over the frequency range 8–26.5 GHz is described which depends on the cavity resonance perturbation method. Equations for the resonance perturbation by a thin capillary in cylindrical as well as in rectangular cavities are derived, demonstrating the dependence of the real and imaginary part of ε on both the resonance frequency shift and the change in resonance width of the cavity. The microwave generator with high frequency stability (1 part in 108) is frequency modulated at one or two frequencies. The use of phase sensitive detection allows ε measurements on water to within 1%, as well as dielectric difference measurements on aqueous solutions to within 0.005 in Δε′ and Δε″ (1 part in 104 or better). Details of cavity constructions are given and some measurements on water and on dilute aqueous solutions are reported.
ISSN:0034-6748
1089-7623
DOI:10.1063/1.1134142