Soft x‐ray microscope at the undulator beamline of the Photon Factory
A soft x‐ray microscope using Fresnel zone plates as optical imaging elements was installed at the Photon Factory. It makes use of undulator radiation as a source of soft x rays ranging from 2.0 to 3.0 nm in wavelength. The optical system was designed to match to the undulator radiation. A performan...
Saved in:
Published in: | Review of Scientific Instruments 1989-07, Vol.60 (7), p.2448-2451 |
---|---|
Main Authors: | , , , , , , |
Format: | Article |
Language: | English |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | A soft x‐ray microscope using Fresnel zone plates as optical imaging elements was installed at the Photon Factory. It makes use of undulator radiation as a source of soft x rays ranging from 2.0 to 3.0 nm in wavelength. The optical system was designed to match to the undulator radiation. A performance test was made at 2.66 nm by various pitches of transmission gratings. Those magnified images were taken with a magnification of 230 and exposure times of about 10 s. The modulation transfer function was measured over the spatial frequency up to 3300 lp/mm. Further, magnified images of some biological specimens were obtained. |
---|---|