Loading…

Installation and initial operation of the Suss Advanced Lithography Model 4 X‐ray Stepper (abstract)

A Suss Advanced Lithography X‐ray Stepper designed as a production tool for high throughput in the sub‐quarter‐micron device range has been installed and is being commissioned at the University of Wisconsin’s Center for X‐ray Lithography (CXrL). Illumination for the stepper is provided by a scanning...

Full description

Saved in:
Bibliographic Details
Published in:Review of scientific instruments 1996-09, Vol.67 (9), p.3358-3358
Main Authors: Wells, Gregory M., Wallace, J. P., Brodsky, E. L., Leonard, Q. J., Reilly, M. T., Anderson, P. D., Lee, W. K., Cerrina, F., Simon, Klaus
Format: Article
Language:English
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:A Suss Advanced Lithography X‐ray Stepper designed as a production tool for high throughput in the sub‐quarter‐micron device range has been installed and is being commissioned at the University of Wisconsin’s Center for X‐ray Lithography (CXrL). Illumination for the stepper is provided by a scanning beamline designed and constructed at CXrL. The beamline optical components are a gold‐coated plane mirror, a 1‐micron‐thick silicon carbide window, and a 25‐micron‐thick beryllium exit window. Beamline features include synchronized scanning of the mirror and exit window, variable scan velocity to compensate for reflectivity changes as a function of incident angle, and a horizontal oscillation of the beryllium window during vertical scanning to average the effects of nonuniform beryllium window transmission. A helium purged snout transports the x‐rays from the beamline exit window, to the exposure plane in the stepper. This snout is retractable to allow for the loading and unloading of masks into the stepper. The motions of the mirror, exit window, and snout are computer controlled by a LABVIEW program that communicates with the stepper control software. The design of the beamline and initial operating experiences with the beamline and stepper will be discussed.
ISSN:0034-6748
1089-7623
DOI:10.1063/1.1147382