The multichannel triple-laser interferometer/polarimeter system at RTP
A 19-channel combined interferometer and polarimeter system has recently become operational at the Rijnhuizen Tokamak Project (a=0.164 m , R=0.72 m , B tor ⩽2.5 T , I p ⩽150 kA , plasma pulse duration ⩽500 ms ) , in order to determine the poloidal magnetic field and plasma density distribution...
Saved in:
Published in: | Review of Scientific Instruments 1997-02, Vol.68 (2), p.1217-1226 |
---|---|
Main Authors: | , , , |
Format: | Article |
Language: | English |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | A 19-channel combined interferometer and polarimeter system has recently become operational at the Rijnhuizen Tokamak Project
(a=0.164
m
,
R=0.72
m
,
B
tor
⩽2.5
T
,
I
p
⩽150
kA
, plasma pulse duration
⩽500
ms
)
, in order to determine the poloidal magnetic field and plasma density distribution simultaneously. The method is based on the separate propagation of two counter-rotating, circularly polarized probing waves, and enables high temporal resolution measurements of both the interferometric and the polarimetric phase change using only one detector per line of sight. In this article a thorough discussion of the (polarization-related) optical, electronic, and computational aspects of the system will be presented. |
---|---|
ISSN: | 0034-6748 1089-7623 |
DOI: | 10.1063/1.1147884 |