A Bakable Changeover Valve for High Precision Mass Spectrometric Comparison of the Isotopic Composition of Gases

The construction of a bakable changeover valve for high precision mass spectrometric comparison of the isotopic composition of gases is described. The leakage across the valve seats is about an order of magnitude smaller than the values published for similar valves in use, and the ``mixing correctio...

Full description

Saved in:
Bibliographic Details
Published in:Review of scientific instruments 1972-01, Vol.43 (10), p.1454-1455
Main Authors: Begbie, P. J., Beckinsale, R. D., Freeman, N. J., Powell, R. E.
Format: Article
Language:English
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:The construction of a bakable changeover valve for high precision mass spectrometric comparison of the isotopic composition of gases is described. The leakage across the valve seats is about an order of magnitude smaller than the values published for similar valves in use, and the ``mixing correction'' which can be a significant correction to the experimental data is negligible.