Loading…
From CCT-K7 to CCT-K7.2021: Approaching the definition of the triple point of water temperature
In 2002–2004, the first CIPM Key Comparison of triple-point-of-water (TPW) cells, CCT-K7, was carried out. One of the major observations in CCT-K7 was the bimodal distribution of the results because 3 out of 21 participating labs had their TPW realization based on Vienna Standard Mean Ocean Water (V...
Saved in:
Main Authors: | , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , |
---|---|
Format: | Conference Proceeding |
Language: | English |
Subjects: | |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | In 2002–2004, the first CIPM Key Comparison of triple-point-of-water (TPW) cells, CCT-K7, was carried out. One of the major observations in CCT-K7 was the bimodal distribution of the results because 3 out of 21 participating labs had their TPW realization based on Vienna Standard Mean Ocean Water (V-SMOW) and thus followed the definition of the TPW given in the Supplementary Information to the International Temperature Scale of 1990 (ITS-90), which resulted in “warmer” national realizations compared to the rest of the participants. Based on the CCT-K7 results, the “Clarification of the definition of the kelvin” was issued in 2005, where the importance of the isotopic corrections to V-SMOW water was emphasized. The successor of CCT-K7, CCT-K7.2021, was carried out in 2021–2022. In this paper, we compare the results of the two comparisons side-by-side to illustrate: a) the improved measurement capabilities of the participants over the past 20 years and b) the improved quality of the TPW cells. As expected, in CCT-K7.2021 most participants renewed their national reference ensembles with newer, higher quality and vitreous-silica cells and applied the isotopic corrections to their respective national references. This, in turn, led to: 1) a smaller spread of the national realizations by almost a factor of two — 28 µK instead of 50 µK in CCT-K7, 2) an only slightly asymmetric distribution of the results, and 3) an increase in KCRV value by approximately 66 µK compared to CCT-K7. |
---|---|
ISSN: | 0094-243X 1551-7616 |
DOI: | 10.1063/5.0234488 |