WE‐D‐BRC‐04: MR‐XRT at 1.5T, the UMC Utrecht Hybrid MRI Linac System

At the UMC Utrecht, the Netherlands we have constructed a prototype MRI accelerator. The design is a 6 MV Elekta (Crawley, U.K.) Compact accelerator combined with a modified 1.5 T Philips Achieva (Best, The Netherlands) MRI system. The systems run according to our specifications. The simultaneous op...

Full description

Saved in:
Bibliographic Details
Published in:Medical Physics 2009-06, Vol.36 (6), p.2775-2775
Main Authors: Lagendijk, J, Raaymakers, B, Overweg, J, Kok, J, Kerkhof, E, Raaijmakers, A, van der Put, R, Meijsing, I, Crijns, S, Benedosso, F, van Vulpen, M, Brown, K
Format: Article
Language:English
Subjects:
Citations: Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
cited_by cdi_FETCH-LOGICAL-c2177-51bfc274895121f357161c6a332a2dcc25b31db797dbfc497a4879fbbdbb395f3
cites
container_end_page 2775
container_issue 6
container_start_page 2775
container_title Medical Physics
container_volume 36
creator Lagendijk, J
Raaymakers, B
Overweg, J
Kok, J
Kerkhof, E
Raaijmakers, A
van der Put, R
Meijsing, I
Crijns, S
Benedosso, F
van Vulpen, M
Brown, K
description At the UMC Utrecht, the Netherlands we have constructed a prototype MRI accelerator. The design is a 6 MV Elekta (Crawley, U.K.) Compact accelerator combined with a modified 1.5 T Philips Achieva (Best, The Netherlands) MRI system. The systems run according to our specifications. The simultaneous operation is shown by performing diagnostic quality 1.5 T MRI with the radiation beam on. As designed the interference between the two systems is completely suppressed, no interference was found. The integrated 1.5 T MRI system and radiotherapy accelerator allow simultaneous irradiation and MR imaging. Both systems operate independent and no synchronization is required. The full diagnostic imaging capacities of the Philips MRI can be used, dedicated sequences for MRI guided radiotherapy treatments will be developed. This proof of concept opens the door towards a clinical prototype to start testing MRI guided Radiotherapy (MRIgRT) in the clinic. Acknowledgements: Both Elekta and Philips participate in this research program
doi_str_mv 10.1118/1.3182527
format article
fullrecord <record><control><sourceid>wiley_scita</sourceid><recordid>TN_cdi_scitation_primary_10_1118_1_3182527</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>MP2527</sourcerecordid><originalsourceid>FETCH-LOGICAL-c2177-51bfc274895121f357161c6a332a2dcc25b31db797dbfc497a4879fbbdbb395f3</originalsourceid><addsrcrecordid>eNp90LtOwzAUBmALgUQpDLyBVxAJPr7UMRuEQislAoVEsEW2k6hBvaDYUpWNR-AZeRKC2hWGcxk-_cOP0DmQEACiawgZRFRQeYBGlEsWcErUIRoRonhAORHH6MS5d0LIhAkyQsnr9Pvz636YuyweNuE3OM2G5y3LsfYYQpFfYb-ocZHGuPBdbRcez3rTtdUA5zhp19ril975enWKjhq9dPXZ_o5R8TDN41mQPD3O49sksBSkDASYxlLJIyWAQsOEhAnYiWaMalpZS4VhUBmpZDVArqTmkVSNMZUxTImGjdHFLtd2G-e6uik_unalu74EUv7WUEK5r2Gwwc5u22Xd_w3L9HnvL3fe2dZr327W_4T_AED6aYc</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype></control><display><type>article</type><title>WE‐D‐BRC‐04: MR‐XRT at 1.5T, the UMC Utrecht Hybrid MRI Linac System</title><source>Wiley-Blackwell Read &amp; Publish Collection</source><creator>Lagendijk, J ; Raaymakers, B ; Overweg, J ; Kok, J ; Kerkhof, E ; Raaijmakers, A ; van der Put, R ; Meijsing, I ; Crijns, S ; Benedosso, F ; van Vulpen, M ; Brown, K</creator><creatorcontrib>Lagendijk, J ; Raaymakers, B ; Overweg, J ; Kok, J ; Kerkhof, E ; Raaijmakers, A ; van der Put, R ; Meijsing, I ; Crijns, S ; Benedosso, F ; van Vulpen, M ; Brown, K</creatorcontrib><description>At the UMC Utrecht, the Netherlands we have constructed a prototype MRI accelerator. The design is a 6 MV Elekta (Crawley, U.K.) Compact accelerator combined with a modified 1.5 T Philips Achieva (Best, The Netherlands) MRI system. The systems run according to our specifications. The simultaneous operation is shown by performing diagnostic quality 1.5 T MRI with the radiation beam on. As designed the interference between the two systems is completely suppressed, no interference was found. The integrated 1.5 T MRI system and radiotherapy accelerator allow simultaneous irradiation and MR imaging. Both systems operate independent and no synchronization is required. The full diagnostic imaging capacities of the Philips MRI can be used, dedicated sequences for MRI guided radiotherapy treatments will be developed. This proof of concept opens the door towards a clinical prototype to start testing MRI guided Radiotherapy (MRIgRT) in the clinic. Acknowledgements: Both Elekta and Philips participate in this research program</description><identifier>ISSN: 0094-2405</identifier><identifier>EISSN: 2473-4209</identifier><identifier>DOI: 10.1118/1.3182527</identifier><identifier>CODEN: MPHYA6</identifier><language>eng</language><publisher>American Association of Physicists in Medicine</publisher><subject>Linear accelerators ; Magnetic resonance imaging ; Medical imaging ; Medical magnetic resonance imaging ; Radiation therapy ; Testing procedures</subject><ispartof>Medical Physics, 2009-06, Vol.36 (6), p.2775-2775</ispartof><rights>American Association of Physicists in Medicine</rights><rights>2009 American Association of Physicists in Medicine</rights><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c2177-51bfc274895121f357161c6a332a2dcc25b31db797dbfc497a4879fbbdbb395f3</citedby></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><link.rule.ids>309,310,314,776,780,785,786,23909,23910,25118,27901,27902</link.rule.ids></links><search><creatorcontrib>Lagendijk, J</creatorcontrib><creatorcontrib>Raaymakers, B</creatorcontrib><creatorcontrib>Overweg, J</creatorcontrib><creatorcontrib>Kok, J</creatorcontrib><creatorcontrib>Kerkhof, E</creatorcontrib><creatorcontrib>Raaijmakers, A</creatorcontrib><creatorcontrib>van der Put, R</creatorcontrib><creatorcontrib>Meijsing, I</creatorcontrib><creatorcontrib>Crijns, S</creatorcontrib><creatorcontrib>Benedosso, F</creatorcontrib><creatorcontrib>van Vulpen, M</creatorcontrib><creatorcontrib>Brown, K</creatorcontrib><title>WE‐D‐BRC‐04: MR‐XRT at 1.5T, the UMC Utrecht Hybrid MRI Linac System</title><title>Medical Physics</title><description>At the UMC Utrecht, the Netherlands we have constructed a prototype MRI accelerator. The design is a 6 MV Elekta (Crawley, U.K.) Compact accelerator combined with a modified 1.5 T Philips Achieva (Best, The Netherlands) MRI system. The systems run according to our specifications. The simultaneous operation is shown by performing diagnostic quality 1.5 T MRI with the radiation beam on. As designed the interference between the two systems is completely suppressed, no interference was found. The integrated 1.5 T MRI system and radiotherapy accelerator allow simultaneous irradiation and MR imaging. Both systems operate independent and no synchronization is required. The full diagnostic imaging capacities of the Philips MRI can be used, dedicated sequences for MRI guided radiotherapy treatments will be developed. This proof of concept opens the door towards a clinical prototype to start testing MRI guided Radiotherapy (MRIgRT) in the clinic. Acknowledgements: Both Elekta and Philips participate in this research program</description><subject>Linear accelerators</subject><subject>Magnetic resonance imaging</subject><subject>Medical imaging</subject><subject>Medical magnetic resonance imaging</subject><subject>Radiation therapy</subject><subject>Testing procedures</subject><issn>0094-2405</issn><issn>2473-4209</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2009</creationdate><recordtype>article</recordtype><recordid>eNp90LtOwzAUBmALgUQpDLyBVxAJPr7UMRuEQislAoVEsEW2k6hBvaDYUpWNR-AZeRKC2hWGcxk-_cOP0DmQEACiawgZRFRQeYBGlEsWcErUIRoRonhAORHH6MS5d0LIhAkyQsnr9Pvz636YuyweNuE3OM2G5y3LsfYYQpFfYb-ocZHGuPBdbRcez3rTtdUA5zhp19ril975enWKjhq9dPXZ_o5R8TDN41mQPD3O49sksBSkDASYxlLJIyWAQsOEhAnYiWaMalpZS4VhUBmpZDVArqTmkVSNMZUxTImGjdHFLtd2G-e6uik_unalu74EUv7WUEK5r2Gwwc5u22Xd_w3L9HnvL3fe2dZr327W_4T_AED6aYc</recordid><startdate>200906</startdate><enddate>200906</enddate><creator>Lagendijk, J</creator><creator>Raaymakers, B</creator><creator>Overweg, J</creator><creator>Kok, J</creator><creator>Kerkhof, E</creator><creator>Raaijmakers, A</creator><creator>van der Put, R</creator><creator>Meijsing, I</creator><creator>Crijns, S</creator><creator>Benedosso, F</creator><creator>van Vulpen, M</creator><creator>Brown, K</creator><general>American Association of Physicists in Medicine</general><scope>AAYXX</scope><scope>CITATION</scope></search><sort><creationdate>200906</creationdate><title>WE‐D‐BRC‐04: MR‐XRT at 1.5T, the UMC Utrecht Hybrid MRI Linac System</title><author>Lagendijk, J ; Raaymakers, B ; Overweg, J ; Kok, J ; Kerkhof, E ; Raaijmakers, A ; van der Put, R ; Meijsing, I ; Crijns, S ; Benedosso, F ; van Vulpen, M ; Brown, K</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c2177-51bfc274895121f357161c6a332a2dcc25b31db797dbfc497a4879fbbdbb395f3</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2009</creationdate><topic>Linear accelerators</topic><topic>Magnetic resonance imaging</topic><topic>Medical imaging</topic><topic>Medical magnetic resonance imaging</topic><topic>Radiation therapy</topic><topic>Testing procedures</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Lagendijk, J</creatorcontrib><creatorcontrib>Raaymakers, B</creatorcontrib><creatorcontrib>Overweg, J</creatorcontrib><creatorcontrib>Kok, J</creatorcontrib><creatorcontrib>Kerkhof, E</creatorcontrib><creatorcontrib>Raaijmakers, A</creatorcontrib><creatorcontrib>van der Put, R</creatorcontrib><creatorcontrib>Meijsing, I</creatorcontrib><creatorcontrib>Crijns, S</creatorcontrib><creatorcontrib>Benedosso, F</creatorcontrib><creatorcontrib>van Vulpen, M</creatorcontrib><creatorcontrib>Brown, K</creatorcontrib><collection>CrossRef</collection><jtitle>Medical Physics</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Lagendijk, J</au><au>Raaymakers, B</au><au>Overweg, J</au><au>Kok, J</au><au>Kerkhof, E</au><au>Raaijmakers, A</au><au>van der Put, R</au><au>Meijsing, I</au><au>Crijns, S</au><au>Benedosso, F</au><au>van Vulpen, M</au><au>Brown, K</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>WE‐D‐BRC‐04: MR‐XRT at 1.5T, the UMC Utrecht Hybrid MRI Linac System</atitle><jtitle>Medical Physics</jtitle><date>2009-06</date><risdate>2009</risdate><volume>36</volume><issue>6</issue><spage>2775</spage><epage>2775</epage><pages>2775-2775</pages><issn>0094-2405</issn><eissn>2473-4209</eissn><coden>MPHYA6</coden><abstract>At the UMC Utrecht, the Netherlands we have constructed a prototype MRI accelerator. The design is a 6 MV Elekta (Crawley, U.K.) Compact accelerator combined with a modified 1.5 T Philips Achieva (Best, The Netherlands) MRI system. The systems run according to our specifications. The simultaneous operation is shown by performing diagnostic quality 1.5 T MRI with the radiation beam on. As designed the interference between the two systems is completely suppressed, no interference was found. The integrated 1.5 T MRI system and radiotherapy accelerator allow simultaneous irradiation and MR imaging. Both systems operate independent and no synchronization is required. The full diagnostic imaging capacities of the Philips MRI can be used, dedicated sequences for MRI guided radiotherapy treatments will be developed. This proof of concept opens the door towards a clinical prototype to start testing MRI guided Radiotherapy (MRIgRT) in the clinic. Acknowledgements: Both Elekta and Philips participate in this research program</abstract><pub>American Association of Physicists in Medicine</pub><doi>10.1118/1.3182527</doi><tpages>1</tpages></addata></record>
fulltext fulltext
identifier ISSN: 0094-2405
ispartof Medical Physics, 2009-06, Vol.36 (6), p.2775-2775
issn 0094-2405
2473-4209
language eng
recordid cdi_scitation_primary_10_1118_1_3182527
source Wiley-Blackwell Read & Publish Collection
subjects Linear accelerators
Magnetic resonance imaging
Medical imaging
Medical magnetic resonance imaging
Radiation therapy
Testing procedures
title WE‐D‐BRC‐04: MR‐XRT at 1.5T, the UMC Utrecht Hybrid MRI Linac System
url http://sfxeu10.hosted.exlibrisgroup.com/loughborough?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-02-05T21%3A30%3A08IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-wiley_scita&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=WE%E2%80%90D%E2%80%90BRC%E2%80%9004:%20MR%E2%80%90XRT%20at%201.5T,%20the%20UMC%20Utrecht%20Hybrid%20MRI%20Linac%20System&rft.jtitle=Medical%20Physics&rft.au=Lagendijk,%20J&rft.date=2009-06&rft.volume=36&rft.issue=6&rft.spage=2775&rft.epage=2775&rft.pages=2775-2775&rft.issn=0094-2405&rft.eissn=2473-4209&rft.coden=MPHYA6&rft_id=info:doi/10.1118/1.3182527&rft_dat=%3Cwiley_scita%3EMP2527%3C/wiley_scita%3E%3Cgrp_id%3Ecdi_FETCH-LOGICAL-c2177-51bfc274895121f357161c6a332a2dcc25b31db797dbfc497a4879fbbdbb395f3%3C/grp_id%3E%3Coa%3E%3C/oa%3E%3Curl%3E%3C/url%3E&rft_id=info:oai/&rft_id=info:pmid/&rfr_iscdi=true