SU‐GG‐I‐08: Optimize Pediatric Abdominal CT Protocols Based on Circumferences

Purpose: There is an increasing demand on lowering the radiation dose in pediatric CT to reduce the long‐term cancer risks. In this study, pediatric CT protocols are optimized based on “equivalent abdominal circumference (EAC)” by varying both kVp and effective mAs (mA*exposure_time/pitch). Methods...

Full description

Saved in:
Bibliographic Details
Published in:Medical Physics 2010-06, Vol.37 (6), p.3102-3103
Main Authors: Dong, F, Davros, W
Format: Article
Language:English
Subjects:
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
cited_by
cites
container_end_page 3103
container_issue 6
container_start_page 3102
container_title Medical Physics
container_volume 37
creator Dong, F
Davros, W
description Purpose: There is an increasing demand on lowering the radiation dose in pediatric CT to reduce the long‐term cancer risks. In this study, pediatric CT protocols are optimized based on “equivalent abdominal circumference (EAC)” by varying both kVp and effective mAs (mA*exposure_time/pitch). Methods and Materials: All measurements were taken using a 16‐row CT scanner. To mimic the shape of the pediatric abdomen, a saline bag was placed on each side of 10cm and 16cm‐diameter, circular cylindrical phantoms. Dose was measured using a 100 mm pencil probe ion chamber. Noise was measured by recording the standard deviation near the center of selected images while kVp, mAs and phantom diameter were varied in turn. Dose was collected at four available kVp settings. At each kVp, measurements were taken at five different mAs. For CNR measurement, iodine contrast was diluted to match clinically relevant CT number of ∼300 HU and this diluted contrast was injected into the center holes of the CTDI phantoms. Distilled water was also injected to one of the peripheral holes as reference. The contrast was measured based on the difference of CT number between iodine and water. Results: Optimal scan techniques are selected for each EAC range based on the clinical noise tolerance < 20HU and CTDIvol < 2.5cGy. For the smallest patient (< 30 cm EAC), scan technique of 80kVp and
doi_str_mv 10.1118/1.3468041
format article
fullrecord <record><control><sourceid>wiley_scita</sourceid><recordid>TN_cdi_scitation_primary_10_1118_1_3468041</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>MP8041</sourcerecordid><originalsourceid>FETCH-LOGICAL-c1771-7bfec6ebefc47c182e027aad06611d42d67b8e26e431979a4b7e6d757eb4dd243</originalsourceid><addsrcrecordid>eNp90LFOwzAUBVALgUQpDPyBV5BSnh3XTthKBKVSUSu1nSPHfpGMkriyg1CZ-AS-kS-hqF1hePctR3e4hFwzGDHGsjs2SoXMQLATMuBCpYngkJ-SAUAuEi5gfE4uYnwFAJmOYUBWq83359d0uo_Z_iC7p4tt71r3gXSJ1uk-OEMnlfWt63RDizVdBt9745tIH3RES31HCxfMW1tjwM5gvCRntW4iXh3_kGyeHtfFczJfTGfFZJ4YphRLVFWjkVhhbYQyLOMIXGltQUrGrOBWqipDLlGkLFe5FpVCadVYYSWs5SIdkptDrwk-xoB1uQ2u1WFXMih_1yhZeVxjb5ODfXcN7v6G5cvy6G8PPhrX69757p_yH1n8bnU</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype></control><display><type>article</type><title>SU‐GG‐I‐08: Optimize Pediatric Abdominal CT Protocols Based on Circumferences</title><source>Wiley</source><creator>Dong, F ; Davros, W</creator><creatorcontrib>Dong, F ; Davros, W</creatorcontrib><description>Purpose: There is an increasing demand on lowering the radiation dose in pediatric CT to reduce the long‐term cancer risks. In this study, pediatric CT protocols are optimized based on “equivalent abdominal circumference (EAC)” by varying both kVp and effective mAs (mA*exposure_time/pitch). Methods and Materials: All measurements were taken using a 16‐row CT scanner. To mimic the shape of the pediatric abdomen, a saline bag was placed on each side of 10cm and 16cm‐diameter, circular cylindrical phantoms. Dose was measured using a 100 mm pencil probe ion chamber. Noise was measured by recording the standard deviation near the center of selected images while kVp, mAs and phantom diameter were varied in turn. Dose was collected at four available kVp settings. At each kVp, measurements were taken at five different mAs. For CNR measurement, iodine contrast was diluted to match clinically relevant CT number of ∼300 HU and this diluted contrast was injected into the center holes of the CTDI phantoms. Distilled water was also injected to one of the peripheral holes as reference. The contrast was measured based on the difference of CT number between iodine and water. Results: Optimal scan techniques are selected for each EAC range based on the clinical noise tolerance &lt; 20HU and CTDIvol &lt; 2.5cGy. For the smallest patient (&lt; 30 cm EAC), scan technique of 80kVp and &lt;100 mAs can be used; for 50 cm EAC, 100 kVp and &lt;60 mAs. The signal‐to‐noise ratio measurement indicated lower kVp can effectively enhance the iodine contrast and increase CNR under the same dose with higher kVp. Conclusions: For pediatric CT, the most effective way to strike the balance between image quality and radiation dose is to adjust dose according to patient abdominal circumference. The EAC‐based technique optimization curve can be used to optimize kVp and mAs.</description><identifier>ISSN: 0094-2405</identifier><identifier>EISSN: 2473-4209</identifier><identifier>DOI: 10.1118/1.3468041</identifier><identifier>CODEN: MPHYA6</identifier><language>eng</language><publisher>American Association of Physicists in Medicine</publisher><subject>Cancer ; Computed tomography ; Dosimetry ; Ionization chambers ; Medical image contrast ; Medical image noise ; Medical image quality ; Medical imaging ; Optimization</subject><ispartof>Medical Physics, 2010-06, Vol.37 (6), p.3102-3103</ispartof><rights>American Association of Physicists in Medicine</rights><rights>2010 American Association of Physicists in Medicine</rights><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><link.rule.ids>309,310,314,780,784,789,790,23930,23931,25140,27924,27925</link.rule.ids></links><search><creatorcontrib>Dong, F</creatorcontrib><creatorcontrib>Davros, W</creatorcontrib><title>SU‐GG‐I‐08: Optimize Pediatric Abdominal CT Protocols Based on Circumferences</title><title>Medical Physics</title><description>Purpose: There is an increasing demand on lowering the radiation dose in pediatric CT to reduce the long‐term cancer risks. In this study, pediatric CT protocols are optimized based on “equivalent abdominal circumference (EAC)” by varying both kVp and effective mAs (mA*exposure_time/pitch). Methods and Materials: All measurements were taken using a 16‐row CT scanner. To mimic the shape of the pediatric abdomen, a saline bag was placed on each side of 10cm and 16cm‐diameter, circular cylindrical phantoms. Dose was measured using a 100 mm pencil probe ion chamber. Noise was measured by recording the standard deviation near the center of selected images while kVp, mAs and phantom diameter were varied in turn. Dose was collected at four available kVp settings. At each kVp, measurements were taken at five different mAs. For CNR measurement, iodine contrast was diluted to match clinically relevant CT number of ∼300 HU and this diluted contrast was injected into the center holes of the CTDI phantoms. Distilled water was also injected to one of the peripheral holes as reference. The contrast was measured based on the difference of CT number between iodine and water. Results: Optimal scan techniques are selected for each EAC range based on the clinical noise tolerance &lt; 20HU and CTDIvol &lt; 2.5cGy. For the smallest patient (&lt; 30 cm EAC), scan technique of 80kVp and &lt;100 mAs can be used; for 50 cm EAC, 100 kVp and &lt;60 mAs. The signal‐to‐noise ratio measurement indicated lower kVp can effectively enhance the iodine contrast and increase CNR under the same dose with higher kVp. Conclusions: For pediatric CT, the most effective way to strike the balance between image quality and radiation dose is to adjust dose according to patient abdominal circumference. The EAC‐based technique optimization curve can be used to optimize kVp and mAs.</description><subject>Cancer</subject><subject>Computed tomography</subject><subject>Dosimetry</subject><subject>Ionization chambers</subject><subject>Medical image contrast</subject><subject>Medical image noise</subject><subject>Medical image quality</subject><subject>Medical imaging</subject><subject>Optimization</subject><issn>0094-2405</issn><issn>2473-4209</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2010</creationdate><recordtype>article</recordtype><recordid>eNp90LFOwzAUBVALgUQpDPyBV5BSnh3XTthKBKVSUSu1nSPHfpGMkriyg1CZ-AS-kS-hqF1hePctR3e4hFwzGDHGsjs2SoXMQLATMuBCpYngkJ-SAUAuEi5gfE4uYnwFAJmOYUBWq83359d0uo_Z_iC7p4tt71r3gXSJ1uk-OEMnlfWt63RDizVdBt9745tIH3RES31HCxfMW1tjwM5gvCRntW4iXh3_kGyeHtfFczJfTGfFZJ4YphRLVFWjkVhhbYQyLOMIXGltQUrGrOBWqipDLlGkLFe5FpVCadVYYSWs5SIdkptDrwk-xoB1uQ2u1WFXMih_1yhZeVxjb5ODfXcN7v6G5cvy6G8PPhrX69757p_yH1n8bnU</recordid><startdate>201006</startdate><enddate>201006</enddate><creator>Dong, F</creator><creator>Davros, W</creator><general>American Association of Physicists in Medicine</general><scope>AAYXX</scope><scope>CITATION</scope></search><sort><creationdate>201006</creationdate><title>SU‐GG‐I‐08: Optimize Pediatric Abdominal CT Protocols Based on Circumferences</title><author>Dong, F ; Davros, W</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c1771-7bfec6ebefc47c182e027aad06611d42d67b8e26e431979a4b7e6d757eb4dd243</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2010</creationdate><topic>Cancer</topic><topic>Computed tomography</topic><topic>Dosimetry</topic><topic>Ionization chambers</topic><topic>Medical image contrast</topic><topic>Medical image noise</topic><topic>Medical image quality</topic><topic>Medical imaging</topic><topic>Optimization</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Dong, F</creatorcontrib><creatorcontrib>Davros, W</creatorcontrib><collection>CrossRef</collection><jtitle>Medical Physics</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Dong, F</au><au>Davros, W</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>SU‐GG‐I‐08: Optimize Pediatric Abdominal CT Protocols Based on Circumferences</atitle><jtitle>Medical Physics</jtitle><date>2010-06</date><risdate>2010</risdate><volume>37</volume><issue>6</issue><spage>3102</spage><epage>3103</epage><pages>3102-3103</pages><issn>0094-2405</issn><eissn>2473-4209</eissn><coden>MPHYA6</coden><abstract>Purpose: There is an increasing demand on lowering the radiation dose in pediatric CT to reduce the long‐term cancer risks. In this study, pediatric CT protocols are optimized based on “equivalent abdominal circumference (EAC)” by varying both kVp and effective mAs (mA*exposure_time/pitch). Methods and Materials: All measurements were taken using a 16‐row CT scanner. To mimic the shape of the pediatric abdomen, a saline bag was placed on each side of 10cm and 16cm‐diameter, circular cylindrical phantoms. Dose was measured using a 100 mm pencil probe ion chamber. Noise was measured by recording the standard deviation near the center of selected images while kVp, mAs and phantom diameter were varied in turn. Dose was collected at four available kVp settings. At each kVp, measurements were taken at five different mAs. For CNR measurement, iodine contrast was diluted to match clinically relevant CT number of ∼300 HU and this diluted contrast was injected into the center holes of the CTDI phantoms. Distilled water was also injected to one of the peripheral holes as reference. The contrast was measured based on the difference of CT number between iodine and water. Results: Optimal scan techniques are selected for each EAC range based on the clinical noise tolerance &lt; 20HU and CTDIvol &lt; 2.5cGy. For the smallest patient (&lt; 30 cm EAC), scan technique of 80kVp and &lt;100 mAs can be used; for 50 cm EAC, 100 kVp and &lt;60 mAs. The signal‐to‐noise ratio measurement indicated lower kVp can effectively enhance the iodine contrast and increase CNR under the same dose with higher kVp. Conclusions: For pediatric CT, the most effective way to strike the balance between image quality and radiation dose is to adjust dose according to patient abdominal circumference. The EAC‐based technique optimization curve can be used to optimize kVp and mAs.</abstract><pub>American Association of Physicists in Medicine</pub><doi>10.1118/1.3468041</doi><tpages>2</tpages></addata></record>
fulltext fulltext
identifier ISSN: 0094-2405
ispartof Medical Physics, 2010-06, Vol.37 (6), p.3102-3103
issn 0094-2405
2473-4209
language eng
recordid cdi_scitation_primary_10_1118_1_3468041
source Wiley
subjects Cancer
Computed tomography
Dosimetry
Ionization chambers
Medical image contrast
Medical image noise
Medical image quality
Medical imaging
Optimization
title SU‐GG‐I‐08: Optimize Pediatric Abdominal CT Protocols Based on Circumferences
url http://sfxeu10.hosted.exlibrisgroup.com/loughborough?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-01-01T17%3A33%3A10IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-wiley_scita&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=SU%E2%80%90GG%E2%80%90I%E2%80%9008:%20Optimize%20Pediatric%20Abdominal%20CT%20Protocols%20Based%20on%20Circumferences&rft.jtitle=Medical%20Physics&rft.au=Dong,%20F&rft.date=2010-06&rft.volume=37&rft.issue=6&rft.spage=3102&rft.epage=3103&rft.pages=3102-3103&rft.issn=0094-2405&rft.eissn=2473-4209&rft.coden=MPHYA6&rft_id=info:doi/10.1118/1.3468041&rft_dat=%3Cwiley_scita%3EMP8041%3C/wiley_scita%3E%3Cgrp_id%3Ecdi_FETCH-LOGICAL-c1771-7bfec6ebefc47c182e027aad06611d42d67b8e26e431979a4b7e6d757eb4dd243%3C/grp_id%3E%3Coa%3E%3C/oa%3E%3Curl%3E%3C/url%3E&rft_id=info:oai/&rft_id=info:pmid/&rfr_iscdi=true