Improved accuracy and extended flow range for a Fleisch pneumotachograph

A large linear flow range and a small instrumental dead space volume are incompatible properties for a pneumotachometer (PTM). The linearity of a Fleisch number 2 PTM is studied for flows up to 6 litre s-1 (nominal range 0-2 litre s-1) with various up- and downstream geometries. It is hypothesised t...

Full description

Saved in:
Bibliographic Details
Published in:Medical & biological engineering & computing 1999-07, Vol.37 (4), p.456-460
Main Authors: Stromberg, NOT, Gronkvist, MJ
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:A large linear flow range and a small instrumental dead space volume are incompatible properties for a pneumotachometer (PTM). The linearity of a Fleisch number 2 PTM is studied for flows up to 6 litre s-1 (nominal range 0-2 litre s-1) with various up- and downstream geometries. It is hypothesised that using an array of calibration factors (conductance; flow/pressure), instead of a single calibration factor over the entire flow range, could improve accuracy and also extend the applicable flow range. The conductance against pressure characteristics are calculated with a previously described weighted averaging technique based on multiple strokes from a precision syringe. A single conductance value gives stroke volume errors in the range of -5 to 3% (0-2 litre s-1) and -6 to 11% (0-6 litre s-1) for validation using the same geometry as for calibration. The pressure dependent conductance improves accuracy to within -3% and 1% independent of flow range. However, for validation using a different geometry than for calibration, errors range from -5% to +8%. The degree of non-linearity varies between the geometries (range 3-15%) and is highest when using a one-directional valve upstream of the PTM and a Y-shaped connector. In conclusion, a pressure-dependent conductance improves accuracy and can also be used to extend the applicable flow range up to at least three times the nominal flow range.
ISSN:0140-0118
1741-0444
1741-0444
DOI:10.1007/BF02513330