Loading…

An Instrument for Measuring Ambient Dose Equivalent, H(10)

The design and calibration of a small and simple instrument for measuring the ambient dose equivalent, H*(10), in photon fields is described. Comprising a thermoluminescence LiF dosemeter inside a 20 mm diameter PMMA sphere, it is capable of measuring the ambient dose equivalent with a nearly isotro...

Full description

Saved in:
Bibliographic Details
Published in:Radiation protection dosimetry 1996-01, Vol.67 (1), p.33-39
Main Authors: Carlsson, C.A., Carlsson, G.A., Lund, E., Matscheko, G., Pettersson, H.B.L.
Format: Article
Language:English
Subjects:
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:The design and calibration of a small and simple instrument for measuring the ambient dose equivalent, H*(10), in photon fields is described. Comprising a thermoluminescence LiF dosemeter inside a 20 mm diameter PMMA sphere, it is capable of measuring the ambient dose equivalent with a nearly isotropic response. In the interval 0.1-100 mSv and for the energy range 30 keV to 1.25 MeV the energy response is within -31% and +15% relative to that of 137Cs gamma radiation (662 keV). In practical use, it is therefore sufficient to calibrate the instrument in a 137Cs gamma field using the corresponding conversion coefficient H*(10)/Kair taken from tabulations. The possibility of using the instrument to monitor the ambient dose equivalent for energies above 1.25 MeV is discussed and indicates that the range of applicability can be extended to 4.4 MeV with an energy response within -10% relative to 662 keV.
ISSN:0144-8420
1742-3406
1742-3406
DOI:10.1093/oxfordjournals.rpd.a031794