Combined measurement of dose and α/γ radiation-field-components using the shape of composite peak 5 in the glow curve of LiF:Mg,Ti
The shape of composite peak 5 in the glow curve of LiF:Mg,Ti (TLD-100), specifically, the ratio of the intensity of peak 5a to peak 5, is employed to separately estimate the gamma and alpha particle components in a mixed radiation field. A blind test demonstrates the reasonable accuracy of the techn...
Saved in:
Published in: | Radiation measurements 2014-12, Vol.71, p.86-89 |
---|---|
Main Authors: | , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | The shape of composite peak 5 in the glow curve of LiF:Mg,Ti (TLD-100), specifically, the ratio of the intensity of peak 5a to peak 5, is employed to separately estimate the gamma and alpha particle components in a mixed radiation field. A blind test demonstrates the reasonable accuracy of the technique which yields results between 3.8% and 30%.
•The combined use of 5a/5 peaks for beta/gamma and alpha particle separation in a mixed radiation field has been demonstrated.•A blind test demonstrated the reasonable accuracy of the technique.•The enhanced relative population of peak 7 to high LET could add further accuracy to these efforts. |
---|---|
ISSN: | 1350-4487 1879-0925 1879-0925 |