Loading…

Simultaneous monitoring of intratidal compliance and resistance in mechanically ventilated piglets: A feasibility study in two different study groups

Highlights • Intratidal resistance–volume profiles change in a characteristic way depending on PEEP, tidal volume and the underlying mechanical properties of the lung. • In healthy lungs, intratidal resistance–volume profiles decline linearly. • A curved intratidal resistance decline might indicate...

Full description

Saved in:
Bibliographic Details
Published in:Respiratory physiology & neurobiology 2015-12, Vol.219, p.36-42
Main Authors: Buehler, Sarah, Schumann, Stefan, Vimláti, László, Lichtwarck-Aschoff, Michael, Guttmann, Josef
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:Highlights • Intratidal resistance–volume profiles change in a characteristic way depending on PEEP, tidal volume and the underlying mechanical properties of the lung. • In healthy lungs, intratidal resistance–volume profiles decline linearly. • A curved intratidal resistance decline might indicate pathological changes in the underlying mechanical lung properties. • Intratidal resistance–volume profiles might help guiding a lung protective ventilation setting.
ISSN:1569-9048
1878-1519
1878-1519
DOI:10.1016/j.resp.2015.08.001