Comparison between Abbott m2000 RealTime and Alinity m STI systems for detection of Chlamydia trachomatis, Neisseria gonorrhoeae, and Mycoplasma genitalium
The new Abbott Alinity m STI Assay was compared with Abbott m2000 RealTime PCR. For Chlamydia trachomatis , 26 (7.5%) of 347 samples were positive in the Alinity assay and 24 (6.9%) in the m2000 assay. Corresponding figures for Neisseria gonorrhoeae were 23 (6.6%) and 17 (4.9%). For Mycoplasma genit...
Saved in:
Published in: | European journal of clinical microbiology & infectious diseases 2021-10, Vol.40 (10), p.2217-2220 |
---|---|
Main Authors: | , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | The new Abbott Alinity m STI Assay was compared with Abbott m2000 RealTime PCR. For
Chlamydia trachomatis
, 26 (7.5%) of 347 samples were positive in the Alinity assay and 24 (6.9%) in the m2000 assay. Corresponding figures for
Neisseria gonorrhoeae
were 23 (6.6%) and 17 (4.9%). For
Mycoplasma genitalium
, 22 (7.9%) of 279 samples were positive in the Alinity assay and 18 (6.5%) in the m2000 assay, for which DNA extraction was performed on an m2000sp instrument combined with in-house real-time PCR. The Alinity assay has at least the same sensitivity as the m2000 assay. The specificity was evaluated by discrepancy analysis. |
---|---|
ISSN: | 0934-9723 1435-4373 1435-4373 |