Loading…

Comparison between Abbott m2000 RealTime and Alinity m STI systems for detection of Chlamydia trachomatis, Neisseria gonorrhoeae, and Mycoplasma genitalium

The new Abbott Alinity m STI Assay was compared with Abbott m2000 RealTime PCR. For Chlamydia trachomatis , 26 (7.5%) of 347 samples were positive in the Alinity assay and 24 (6.9%) in the m2000 assay. Corresponding figures for Neisseria gonorrhoeae were 23 (6.6%) and 17 (4.9%). For Mycoplasma genit...

Full description

Saved in:
Bibliographic Details
Published in:European journal of clinical microbiology & infectious diseases 2021-10, Vol.40 (10), p.2217-2220
Main Authors: Herrmann, Björn, Malm, Karin
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:The new Abbott Alinity m STI Assay was compared with Abbott m2000 RealTime PCR. For Chlamydia trachomatis , 26 (7.5%) of 347 samples were positive in the Alinity assay and 24 (6.9%) in the m2000 assay. Corresponding figures for Neisseria gonorrhoeae were 23 (6.6%) and 17 (4.9%). For Mycoplasma genitalium , 22 (7.9%) of 279 samples were positive in the Alinity assay and 18 (6.5%) in the m2000 assay, for which DNA extraction was performed on an m2000sp instrument combined with in-house real-time PCR. The Alinity assay has at least the same sensitivity as the m2000 assay. The specificity was evaluated by discrepancy analysis.
ISSN:0934-9723
1435-4373
1435-4373
DOI:10.1007/s10096-020-04135-9