Simultaneous determination of fluorine and iodine in urine by ion chromatography with electrochemical pretreatment
A new method for the simultaneous determination of fluorine and iodine in urine by ion chromatography (IC) with electrochemical pretreatment has been developed. The pretreatment was performed in a novel electrochemical oxidationneutralization device (EOND), in which iodide of the sample was oxidized...
Saved in:
Published in: | Chinese chemical letters 2009-12, Vol.20 (12), p.1483-1486 |
---|---|
Main Authors: | , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | A new method for the simultaneous determination of fluorine and iodine in urine by ion chromatography (IC) with electrochemical pretreatment has been developed. The pretreatment was performed in a novel electrochemical oxidationneutralization device (EOND), in which iodide of the sample was oxidized to iodate and the alkaline digestion sample solution was neutralized. Under the optimized conditions, the limits of detection (LOD, S/N = 3) were 2.5 μg/L for fluoride and 20 μg/L for iodate, respectively. The recoveries were in the range of 93-102% for fluoride and 86-98% for iodate. |
---|---|
ISSN: | 1001-8417 1878-5964 |