An improved method for flat-field correction of flat panel x-ray detector

In this Technical Note, the effects of different flat-field techniques are examined for a cesium iodide flat panel detector, which exhibited a slightly nonlinear exposure response. The results indicate that the variable flat-field correction method with the appropriate polynomial fit provides excell...

Full description

Saved in:
Bibliographic Details
Published in:Medical physics (Lancaster) 2006-02, Vol.33 (2), p.391-393
Main Authors: Kwan, Alexander L. C., Seibert, J. Anthony, Boone, John M.
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:In this Technical Note, the effects of different flat-field techniques are examined for a cesium iodide flat panel detector, which exhibited a slightly nonlinear exposure response. The results indicate that the variable flat-field correction method with the appropriate polynomial fit provides excellent correction throughout the entire exposure range. The averaged normalized variation factor, used to assess the nonuniformity of the flat-field correction, decreased from 30.76 for the fixed correction method to 4.13 for the variable flat-field correction method with a fourth-order polynomial fit for the 60 kVp spectrum, and from 16.42 to 3.97 for the 95 kVp spectrum.