Evaluation of a commercial three-dimensional electron beam treatment planning system

We evaluated a commercial three-dimensional (3D) electron beam treatment planning system (CADPLAN V.2.7.9) using both experimentally measured and Monte Carlo calculated dose distributions to compare with those predicted by CADPLAN calculations. Tests were carried out at various field sizes and elect...

Full description

Saved in:
Bibliographic Details
Published in:Medical physics (Lancaster) 1999-12, Vol.26 (12), p.2571-2580
Main Authors: Ding, G. X., Cygler, J. E., Zhang, G. G., Yu, M. K.
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:We evaluated a commercial three-dimensional (3D) electron beam treatment planning system (CADPLAN V.2.7.9) using both experimentally measured and Monte Carlo calculated dose distributions to compare with those predicted by CADPLAN calculations. Tests were carried out at various field sizes and electron beam energies from 6 to 20 MeV. For a homogeneous water phantom the agreement between measured and CADPLAN calculated dose distributions is very good except at the phantom surface. CADPLAN is able to predict hot and cold spots caused by a simple 3D inhomogeneity but unable to predict dose distributions for a more complex geometry where CADPLAN underestimates dose changes caused by inhomogeneity. We discussed possible causes for the inaccuracy in the CADPLAN dose calculations. In addition, we have tested CADPLAN treatment monitor unit and electron cut-out factor calculations and found that CADPLAN predictions generally agree with manual calculations.
ISSN:0094-2405
2473-4209
DOI:10.1118/1.598795