Experimental investigation of the characteristics of radioactive beams for heavy ion therapy
Purpose This work has two related objectives. The first is to estimate the relative biological effectiveness of two radioactive heavy ion beams based on experimental measurements, and compare these to the relative biological effectiveness of corresponding stable isotopes to determine whether they ar...
Saved in:
Published in: | Medical physics (Lancaster) 2020-07, Vol.47 (7), p.3123-3132 |
---|---|
Main Authors: | , , , , , , , , , , , , , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Purpose
This work has two related objectives. The first is to estimate the relative biological effectiveness of two radioactive heavy ion beams based on experimental measurements, and compare these to the relative biological effectiveness of corresponding stable isotopes to determine whether they are therapeutically equivalent. The second aim is to quantitatively compare the quality of images acquired postirradiation using an in‐beam whole‐body positron emission tomography scanner for range verification quality assurance.
Methods
The energy deposited by monoenergetic beams of 11C at 350 MeV/u, 15O at 250 MeV/u, 12C at 350 MeV/u, and 16O at 430 MeV/u was measured using a cruciform transmission ionization chamber in a water phantom at the Heavy Ion Medical Accelerator in Chiba (HIMAC), Japan. Dose‐mean lineal energy was measured at various depths along the path of each beam in a water phantom using a silicon‐on‐insulator mushroom microdosimeter. Using the modified microdosimetric kinetic model, the relative biological effectiveness at 10% survival fraction of the radioactive ion beams was evaluated and compared to that of the corresponding stable ions along the path of the beam. Finally, the postirradiation distributions of positron annihilations resulting from the decay of positron‐emitting nuclei were measured for each beam in a gelatin phantom using the in‐beam whole‐body positron emission tomography scanner at HIMAC. The depth of maximum positron‐annihilation density was compared with the depth of maximum dose deposition and the signal‐to‐background ratios were calculated and compared for images acquired over 5 and 20 min postirradiation of the phantom.
Results
In the entrance region, the hboxRBE10 was 1.2 ± 0.1 for both 11C and 12C beams, while for 15O and 16O it was 1.4 ± 0.1 and 1.3 ± 0.1, respectively. At the Bragg peak, the RBE10 was 2.7 ± 0.4 for 11C and 2.9 ± 0.4 for 12C, while for 15O and 16O it was 2.7 ± 0.4 and 2.8 ± 0.4, respectively. In the tail region, RBE10 could only be evaluated for carbon; the RBE10 was 1.6 ± 0.2 and 1.5 ± 0.1 for 11C and 12C, respectively. Positron emission tomography images obtained from gelatin targets irradiated by radioactive ion beams exhibit markedly improved signal‐to‐background ratios compared to those obtained from targets irradiated by nonradioactive ion beams, with 5‐fold and 11‐fold increases in the ratios calculated for the 15O and 11C images compared with the values obtained for 16O and 12C, respectively. |
---|---|
ISSN: | 0094-2405 2473-4209 |
DOI: | 10.1002/mp.14177 |