Loading…
Development of head docking device for linac-based radiosurgery with a Neptun 10 PC linac
Stereotactic radiosurgery is a method for focused irradiation of intracranial lesions. Linac-based radiosurgery is currently performed by two techniques: couch mounted and pedestal mounted. In the first technique a device is required to affix the patient's head to the couch and neoreover to tra...
Saved in:
Published in: | Physica medica 2006, Vol.22 (1), p.25-28 |
---|---|
Main Authors: | , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Stereotactic radiosurgery is a method for focused irradiation of intracranial lesions. Linac-based radiosurgery is currently performed by two techniques: couch mounted and pedestal mounted. In the first technique a device is required to affix the patient's head to the couch and neoreover to translate it accurately. Structure of such a device constructed by the authors plus acceptance test performed for evaluation is described in the article.
A head docking device has been designed and constructed according to geometry of linac's couch and also desired functions. The device is cornpletely made from aluminum and consists of four major components: attachment bar, lower structure with four moveing accuracy mechanical stability and isocentric accuracy were assessed in the frame of acceptance test.
Translating accuracy, mechanical stability and isocentric accuracy were found to be respectively: 1 mm, 1.64 mm and 3.2 mm with accuracy of 95%.
According to AAPM report no. 54, a head docking device should translate head with an accuracy of 1 mm; this recommendation has been met. Moreover, we have demonstrated that the isocentric accuracy and mechanical stability of the device are sufficient that the device on confidently be used in stereotactic treatment. |
---|---|