Loading…

Determining the need for repeat testing of blood ethanol concentration: Evaluation of the synchron blood ethyl alcohol assay kit

In clinical laboratories, a common practice used to verify tests prior to reporting is repeat testing. Our objective was to evaluate the differences between the results of blood ethanol concentration (BEC) test repetitions and report on the role of repeat testing to prevent reporting of incorrect re...

Full description

Saved in:
Bibliographic Details
Published in:Journal of medical biochemistry 2019-07, Vol.38 (3), p.317-322
Main Authors: Ustundağ, Yasemin, Huysal, Kağan, Eren, Sevim Eşmedere
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:In clinical laboratories, a common practice used to verify tests prior to reporting is repeat testing. Our objective was to evaluate the differences between the results of blood ethanol concentration (BEC) test repetitions and report on the role of repeat testing to prevent reporting of incorrect results. We conducted a retrospective study of data retrieved from the Bursa Yuksek Ihtisas Training and Research Hospital's document management system by calculating the percentage change between repeated BEC test runs. To assess for clinical relevance, the bias between two results from the same sample was compared using the 1988 Clinical Laboratory Improvement Amendments' (CLIA) proficiency testing allowable total error (TEa) limits. From a total of 1,627 BEC tests performed between January 2017 and January 2018, 70% (1,133) were repeat tested. Of these, 830 resulted in BECs between 0-5 mmol/L, of which 237 (28.5%) were above the 25% acceptable TEa. Two hundred seventy-six BEC test results were greater than >14 mmol/L, and there was a good consensus between the initial and repeat test results (99%). In this group, the mean bias was 0.0% (95%, CI = -9.8-9.8%). However, three of the repeat test results were considered significantly different. There were two discordant results in the 5-14 mmol/L ethanol level, and the mean bias was 2.1% (95%, CI = -15.0-19.1%). The majority of the repeated BEC test values were the same as the baseline value; therefore, there may be limited benefit in continuing such frequent repeated analyses.
ISSN:1452-8258
1452-8266
1452-8266
DOI:10.2478/jomb-2018-0032