Thermal Comfort and Safety of Cotton Blankets Warmed at 130°F and 200°F

In 2009, the ECRI Institute recommended warming cotton blankets in cabinets set at 130°F or less. However, there is limited research to support the use of this cabinet temperature. To measure skin temperatures and thermal comfort in healthy volunteers before and after application of blankets warmed...

Full description

Saved in:
Bibliographic Details
Published in:Journal of perianesthesia nursing 2013-12, Vol.28 (6), p.337-346
Main Authors: Kelly, Patricia A., Cooper, Susan K., Krogh, Mary L., Morse, Elizabeth C., Crandall, Craig G., Winslow, Elizabeth H., Balluck, Julie P.
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:In 2009, the ECRI Institute recommended warming cotton blankets in cabinets set at 130°F or less. However, there is limited research to support the use of this cabinet temperature. To measure skin temperatures and thermal comfort in healthy volunteers before and after application of blankets warmed in cabinets set at 130 and 200°F, respectively, and to determine the time-dependent cooling of cotton blankets after removal from warming cabinets set at the two temperatures. Prospective, comparative, descriptive. Participants (n = 20) received one or two blankets warmed in 130 or 200°F cabinets. First, skin temperatures were measured, and thermal comfort reports were obtained at fixed timed intervals. Second, blanket temperatures (n = 10) were measured at fixed intervals after removal from the cabinets. No skin temperatures approached levels reported in the literature that cause epidermal damage. Thermal comfort reports supported using blankets from the 200°F cabinet, and blankets lost heat quickly over time. We recommend warming cotton blankets in cabinets set at 200°F or less to improve thermal comfort without compromising patient safety.
ISSN:1089-9472
1532-8473
DOI:10.1016/j.jopan.2012.10.007