The Use of Sound Level Meter Apps in the Clinical Setting
The purpose of this study was to compare sound level meter (SLM) readings obtained using a Larson-Davis (Depew, NY) Model 831 Type 1 SLM, a RadioShack (Fort Worth, TX) SLM, and iPhone 5 (Apple, Cupertino, CA) SLM apps. In Procedure 1, pure tones were measured in an anechoic chamber (125, 250, 500, 1...
Saved in:
Published in: | American journal of speech-language pathology 2016-02, Vol.25 (1), p.14-28 |
---|---|
Main Authors: | , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | The purpose of this study was to compare sound level meter (SLM) readings obtained using a Larson-Davis (Depew, NY) Model 831 Type 1 SLM, a RadioShack (Fort Worth, TX) SLM, and iPhone 5 (Apple, Cupertino, CA) SLM apps.
In Procedure 1, pure tones were measured in an anechoic chamber (125, 250, 500, 1000, 2000, 4000, and 8000 Hz); sound pressure levels (SPLs) ranged from 60 to 100 dB SPL in 10-dB increments. In Procedure 2, human voices were measured. Participants were 20 vocally healthy adults (7 women, 13 men; mean age = 25.1 years). The task was to sustain a vowel "ah" at 3 intensity levels: soft, habitual, and loud. Microphones were lined up equal distances from the participant's mouth, and recordings were captured simultaneously.
Overall, the 3 SLM apps and the RadioShack SLM yielded inconsistent readings compared with the Type 1 SLM.
The use of apps for SPL readings in the clinical setting is premature because all 3 apps adopted were incomparable with the Type 1 SLM. |
---|---|