Development, Preclinical Validation, and Clinical Translation of a Cardiac Magnetic Resonance - Electrophysiology System With Active Catheter Tracking for Ablation of Cardiac Arrhythmia
This study sought to develop an actively tracked cardiac magnetic resonance-guided electrophysiology (CMR-EP) system and perform first-in-human clinical ablation procedures. CMR-EP offers high-resolution anatomy, arrhythmia substrate, and ablation lesion visualization in the absence of ionizing radi...
Saved in:
Published in: | JACC. Clinical electrophysiology 2017-02, Vol.3 (2), p.89-103 |
---|---|
Main Authors: | , , , , , , , , , , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | This study sought to develop an actively tracked cardiac magnetic resonance-guided electrophysiology (CMR-EP) system and perform first-in-human clinical ablation procedures.
CMR-EP offers high-resolution anatomy, arrhythmia substrate, and ablation lesion visualization in the absence of ionizing radiation. Implementation of active tracking, where catheter position is continuously transmitted in a manner analogous to electroanatomic mapping (EAM), is crucial for CMR-EP to take the step from theoretical technology to practical clinical tool.
The setup integrated a clinical 1.5-T scanner, an EP recording and ablation system, and a real-time image guidance platform with components undergoing ex vivo validation. The full system was assessed using a preclinical study (5 pigs), including mapping and ablation with histological validation. For the clinical study, 10 human subjects with typical atrial flutter (age 62 ± 15 years) underwent MR-guided cavotricuspid isthmus (CTI) ablation.
The components of the CMR-EP system were safe (magnetically induced torque, radiofrequency heating) and effective in the CMR environment (location precision). Targeted radiofrequency ablation was performed in all animals and 9 (90%) humans. Seven patients had CTI ablation completed using CMR guidance alone; 2 patients required completion under fluoroscopy, with 2 late flutter recurrences. Acute and chronic CMR imaging demonstrated efficacious lesion formation, verified with histology in animals. Anatomic shape of the CTI was an independent predictor of procedural success.
CMR-EP using active catheter tracking is safe and feasible. The CMR-EP setup provides an effective workflow and has the potential to change the way in which ablation procedures may be performed.