CT automated exposure control using a generalized detectability index
Purpose Identifying an appropriate tube current setting can be challenging when using iterative reconstruction due to the varying relationship between spatial resolution, contrast, noise, and dose across different algorithms. This study developed and investigated the application of a generalized det...
Saved in:
Published in: | Medical physics (Lancaster) 2019-01, Vol.46 (1), p.140-151 |
---|---|
Main Authors: | , , , , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Purpose
Identifying an appropriate tube current setting can be challenging when using iterative reconstruction due to the varying relationship between spatial resolution, contrast, noise, and dose across different algorithms. This study developed and investigated the application of a generalized detectability index (dgen′) to determine the noise parameter to input to existing automated exposure control (AEC) systems to provide consistent image quality (IQ) across different reconstruction approaches.
Methods
This study proposes a task‐based automated exposure control (AEC) method using a generalized detectability index (dgen′). The proposed method leverages existing AEC methods that are based on a prescribed noise level. The generalized dgen′ metric is calculated using lookup tables of task‐based modulation transfer function (MTF) and noise power spectrum (NPS). To generate the lookup tables, the American College of Radiology CT accreditation phantom was scanned on a multidetector CT scanner (Revolution CT, GE Healthcare) at 120 kV and tube current varied manually from 20 to 240 mAs. Images were reconstructed using a reference reconstruction algorithm and four levels of an in‐house iterative reconstruction algorithm with different regularization strengths (IR1–IR4). The task‐based MTF and NPS were estimated from the measured images to create lookup tables of scaling factors that convert between dgen′ and noise standard deviation. The performance of the proposed dgen′‐AEC method in providing a desired IQ level over a range of iterative reconstruction algorithms was evaluated using the American College of Radiology (ACR) phantom with elliptical shell and using a human reader evaluation on anthropomorphic phantom images.
Results
The study of the ACR phantom with elliptical shell demonstrated reasonable agreement between the dgen′ predicted by the lookup table and d′ measured in the images, with a mean absolute error of 15% across all dose levels and maximum error of 45% at the lowest dose level with the elliptical shell. For the anthropomorphic phantom study, the mean reader scores for images resulting from the dgen′‐AEC method were 3.3 (reference image), 3.5 (IR1), 3.6 (IR2), 3.5 (IR3), and 2.2 (IR4). When using the dgen′‐AEC method, the observers’ IQ scores for the reference reconstruction were statistical equivalent to the scores for IR1, IR2, and IR3 iterative reconstructions (P > 0.35). The dgen′‐AEC method achieved this equivalent IQ at lower dose for th |
---|---|
ISSN: | 0094-2405 2473-4209 |
DOI: | 10.1002/mp.13286 |