Loading…

Evaluation of a New Generation Automated Assay for 25-Hydroxy Vitamin D Based on Competitive Protein Binding

BACKGROUNDThe interest for vitamin D has exponentially increased testing demand for 25-hydroxy vitamin D [25(OH)D]. Consequently, many laboratories are switching from LC-MS/MS methods to automated, high-throughput immunoassays. One of the major potential issues with these assays has been the lack of...

Full description

Saved in:
Bibliographic Details
Published in:The journal of applied laboratory medicine 2019-09, Vol.4 (2), p.247-253
Main Authors: Asif, Maryam, Groboske, Sarah E, Leung, Edward K Y, Yeo, Kiang-Teck J, van Wijk, Xander M R
Format: Article
Language:English
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:BACKGROUNDThe interest for vitamin D has exponentially increased testing demand for 25-hydroxy vitamin D [25(OH)D]. Consequently, many laboratories are switching from LC-MS/MS methods to automated, high-throughput immunoassays. One of the major potential issues with these assays has been the lack of cross-reactivity with 25(OH)D2. METHODSWe have evaluated the Roche Elecsys vitamin D total II assay for accuracy by comparing 79 patient samples with LC-MS/MS. The cross-reactivity for 25(OH)D2 was evaluated by analyzing samples with high 25(OH)D2 separately and estimating 25(OH)D2 recovery, as well as by spiking of 25(OH)D2. The assay was further evaluated for precision, linearity, sample type, and common interferences. RESULTSThere was mostly good agreement between the Elecsys and LC-MS/MS assays (Deming regression: y = 0.95x + 0.70), with an overall bias of 2.3% (-0.84 ng/mL). However, there were 6 out of 79 (7.6%) discordant samples. The Deming regression for samples with high 25(OH)D2 compared to LC-MS/MS showed similar slope and intercept (y = 0.97x - 1.1). The average recovery of 25(OH)D2 for these samples was 90%. The initial precision studies were in general agreement with the package insert, but long-term clinical use showed higher-than-claimed imprecision (11.7%-14.4% at 12 ng/mL and 6.9%-7.6% at 27 ng/mL; claimed: 7.2% and 5.0%, respectively). We observed 1 falsely high result in plasma, an issue previously addressed by Roche in a medical device correction. CONCLUSIONSThe analytical performance of the Roche Vitamin D assay was acceptable, and the assay had a good cross-reactivity for 25(OH)D2.
ISSN:2576-9456
2475-7241
DOI:10.1373/jalm.2018.028415