Loading…
Run‐time motion and first‐order shim control by expanded servo navigation
Purpose To provide a navigator‐based run‐time motion and first‐order field correction for three‐dimensional human brain imaging with high precision, minimal calibration and acquisition, and fast processing. Methods A complex‐valued linear perturbation model with feedback control is extended to estim...
Saved in:
Published in: | Magnetic resonance in medicine 2025-01, Vol.93 (1), p.166-182 |
---|---|
Main Authors: | , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Purpose
To provide a navigator‐based run‐time motion and first‐order field correction for three‐dimensional human brain imaging with high precision, minimal calibration and acquisition, and fast processing.
Methods
A complex‐valued linear perturbation model with feedback control is extended to estimate and correct for gradient shim fields using orbital navigators (2.3 ms). Two approaches for sensitizing the model to gradient fields are presented, one based on finite differences with three additional navigators, and another projection‐based approximation requiring no additional navigators. A mechanism for noise decorrelation of the matrix and the data is proposed and evaluated to reduce unwanted parameter biases.
Results
The rigid motion and first‐order field control achieves robust motion and gradient shim corrections improving image quality in a series of phantom and in vivo experiments with varying field conditions. In phantom scans, magnet drifts, forced gradient field perturbations and field distortions from shifts of a second bottle phantom are successfully corrected. Field estimates of the magnet drifts are in good agreement with concurrent field probe measurements. For in vivo scans, the proposed method mitigates field variations from torso motions while being robust to head motion. In vivo gradient field precisions were 30 nT/m$$ 30\;\mathrm{nT}/\mathrm{m} $$ along with single‐digit micrometer and millidegree rigid precisions.
Conclusion
The navigator‐based method achieves accurate, high‐precision run‐time motion and field corrections with low sequence impact and calibration requirements. |
---|---|
ISSN: | 0740-3194 1522-2594 1522-2594 |
DOI: | 10.1002/mrm.30262 |