Loading…
Experimental Study of Products Based on Biocompatible Polymer Material from Methacrylic Oligomers as a Potential Barrier for Preventing Adhesions in Cardiac Surgery
The effectiveness and safety of two types of samples based on a biocompatible polymer material made of methacrylic oligomers (Reperen) as a potential antiadhesion pericardial barrier were evaluated in in vitro and in vivo experiments. Two kinds of samples, reinforced with a polyamide mesh and withou...
Saved in:
Published in: | Bulletin of experimental biology and medicine 2024-11, Vol.178 (1), p.163-168 |
---|---|
Main Authors: | , , , , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | The effectiveness and safety of two types of samples based on a biocompatible polymer material made of methacrylic oligomers (Reperen) as a potential antiadhesion pericardial barrier were evaluated in
in vitro
and
in vivo
experiments. Two kinds of samples, reinforced with a polyamide mesh and without reinforcement, were used. In
in vitro
experiments, no adhesion and aggregation of human fibroblasts to the test samples were detected. In
in vivo
experiments, the samples implanted to rats into the thigh muscles were easily separated from the surrounding tissues 1, 2, and 3 weeks after implantation, being weakly fixed only in the area of the edges. Histological examination at week 2 after implantation revealed no differences between the experimental and control groups. At week 1 and 3, fibrosis and inflammation were more pronounced in animals of the control group (with simulated implantation). The properties demonstrated by both samples of Reperen barriers (with and without polyamide mesh reinforcement)
in vivo
and
in vitro
allow considering them as a potential antiadhesion pericardial barrier for clinical use. |
---|---|
ISSN: | 0007-4888 1573-8221 1573-8221 |
DOI: | 10.1007/s10517-024-06300-y |