Loading…
Quantitative determination of vigabatrin and gabapentin in human serum by gas chromatography–mass spectrometry
Published methods for routine clinical monitoring of vigabatrin and gabapentin are often very laborious. A simple GC-MS method was developed for the simultaneous quantitative determination of vigabatrin and gabapentin in human serum. After protein precipitation, the compounds are derivatized by meth...
Saved in:
Published in: | Clinica chimica acta 2005-04, Vol.354 (1), p.147-151 |
---|---|
Main Authors: | , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Published methods for routine clinical monitoring of vigabatrin and gabapentin are often very laborious. A simple GC-MS method was developed for the simultaneous quantitative determination of vigabatrin and gabapentin in human serum.
After protein precipitation, the compounds are derivatized by methylation and analysed on a polydimethylsiloxane column using splitless injection. Cyclobarbital is used as the internal standard. To attain maximal sensitivity, detection is performed in selected ion monitoring mode.
The method was fully validated and linear calibration curves were obtained in the concentration ranges from 5 to 80 μg/mL for vigabatrin and from 5 to 30 μg/mL for gabapentin. The within-day and day-to-day relative standard deviations at three different concentration levels were |
---|---|
ISSN: | 0009-8981 1873-3492 |
DOI: | 10.1016/j.cccn.2004.11.023 |