Comparison of test devices for skin prick testing

Allergy skin testing guides developing avoidance plans and writing an immunotherapy prescription. The goal for the allergist is to apply allergen skin testing to the appropriate patient population by using a device that minimizes both false-negative and false-positive findings while minimizing patie...

Full description

Saved in:
Bibliographic Details
Published in:Journal of allergy and clinical immunology 2005-08, Vol.116 (2), p.341-346
Main Authors: Carr, Warner W., Martin, Bryan, Howard, Robin S., Cox, Linda, Borish, Larry, the Immunotherapy Committee of the American Academy of Allergy, Asthma and Immunology
Format: Article
Language:English
Subjects:
Citations: Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:Allergy skin testing guides developing avoidance plans and writing an immunotherapy prescription. The goal for the allergist is to apply allergen skin testing to the appropriate patient population by using a device that minimizes both false-negative and false-positive findings while minimizing patient discomfort. New skin testing devices continue to be developed with a trend toward production of multiheaded devices. Data on the performance of these devices in a head-to-head prospective fashion are limited. Our goal was to study 8 commonly used devices to compare their performance in a head-to-head fashion. In a prospective, double-blind fashion, the performance of 8 skin test devices was evaluated. Devices were tested with histamine and saline on both the arms and back of each subject. Devices were rotated over 4 testing sessions, at least a week apart, so each device was tested in each anatomic testing location. Performance elements examined included wheal, flare, pain, sensitivity, specificity, and intradevice variability. We found significant differences in all areas of device performance among all devices examined. Multiheaded devices also demonstrated significant intradevice variability and were more painful than single devices. Furthermore, multiheaded devices had larger reactions on the back, whereas single devices had larger reactions on the arms. Statistically significant differences exist among all devices tested. Providers should consider this data when choosing a device that suits their practice setting and ensure that technicians are sufficiently trained on the correct use of that device.
ISSN:0091-6749
1097-6825
DOI:10.1016/j.jaci.2005.03.035