A comparison of anaesthetic tensions in arterial blood and oxygenator exhaust gas during cardiopulmonary bypass
This study evaluates the usefulness of the analysis of gas sampled from the exhaust port of a membrane oxygenator in the estimation of anaesthetic tension in arterial blood. Sixty‐seven arterial blood samples were drawn from patients undergoing hypothermic cardiopulmonary bypass with anaesthesia mai...
Saved in:
Published in: | Anaesthesia 1999-05, Vol.54 (5), p.434-436 |
---|---|
Main Authors: | , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | This study evaluates the usefulness of the analysis of gas sampled from the exhaust port of a membrane oxygenator in the estimation of anaesthetic tension in arterial blood. Sixty‐seven arterial blood samples were drawn from patients undergoing hypothermic cardiopulmonary bypass with anaesthesia maintained by either isoflurane or desflurane. Anaesthetic tensions in the oxygenator exhaust gas were measured using an infrared analyser and in arterial blood using a two‐stage headspace technique with a gas chromatograph. Both measurement systems were calibrated with the same standard gas mixtures. There was no difference in anaesthetic tension measured in arterial blood and gas leaving the oxygenator exhaust (isoflurane: n = 29, range: 0.3–0.8%, 95% limits of agreement: − 0.08% to 0.09%; desflurane: n = 38, range: 1.5–5.4%; 95% limits of agreement − 0.65% to 0.58%). We conclude that anaesthetic tensions in arterial blood can be accurately monitored by analysis of the gas emerging from the exhaust port of a membrane oxygenator. |
---|---|
ISSN: | 0003-2409 1365-2044 |
DOI: | 10.1046/j.1365-2044.1999.00847.x |