Loading…

“Accordion” Deformity of a Tortuous External Iliac Artery after Stent-Graft Placement

Purpose: To identify a complication of endograft deployment in aneurysmal iliac arteries. Case Report: A 71-year-old man was referred for endovascular treatment of a 60-mm diameter right common iliac artery aneurysm; however, deployment of a homemade covered stent (Palmaz-Schatz and polytetrafluoroe...

Full description

Saved in:
Bibliographic Details
Published in:Journal of endovascular therapy 2001-02, Vol.8 (1), p.93-98
Main Authors: Quinn, Stephen F., Kim, Jinho, Sheley, Robert C., Frankhouse, Joseph H.
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:Purpose: To identify a complication of endograft deployment in aneurysmal iliac arteries. Case Report: A 71-year-old man was referred for endovascular treatment of a 60-mm diameter right common iliac artery aneurysm; however, deployment of a homemade covered stent (Palmaz-Schatz and polytetrafluoroethylene) induced shortening of the tortuous external iliac artery, causing an “accordion” deformity. The anomaly proved difficult to treat with serial Wallstent deployment, because the convolution tightened and migrated caudally with each stent deployed, threatening outflow. Finally, after 3 Wallstents were implanted, the contour of the external iliac artery was straight, and flow was unimpeded. However, 3 weeks later, the external iliac artery had recoiled to its original redundant appearance, but flow remained satisfactory. The aneurysm remains excluded, with satisfactory distal flow after 24 months. Conclusions: Implanting endografts in redundant, tortuous arterial segments may prove problematic, since induced straightening by the device precipitates kinking in the redundant system. Although treatment may be required in some situations, the vessels may return to a noncompressed state by removing the delivery system and guidewire.
ISSN:1526-6028
1545-1550
DOI:10.1177/152660280100800116