Determination of cardiac output in neonatal foals by ultrasound velocity dilution and its comparison to the lithium dilution method
Objective – To compare cardiac output (CO) measured by use of lithium dilution (LiDCO) and ultrasound velocity dilution (UDCO) in conditions of high, intermediate, and low CO in anesthetized foals. Design – Original prospective study. Setting – University teaching hospital. Animals – Six foals 1–3 d...
Saved in:
Published in: | Journal of veterinary emergency and critical care (San Antonio, Tex. : 2000) Tex. : 2000), 2009-10, Vol.19 (5), p.438-443 |
---|---|
Main Authors: | , , , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Objective
–
To compare cardiac output (CO) measured by use of lithium dilution (LiDCO) and ultrasound velocity dilution (UDCO) in conditions of high, intermediate, and low CO in anesthetized foals.
Design
–
Original prospective study.
Setting
–
University teaching hospital.
Animals
–
Six foals 1–3 days of age (38–45 kg).
Interventions
–
Neonatal foals were anesthetized and instrumented to measure direct blood pressure, heart rate, arterial blood gases, and CO. The CO was measured by use of LiDCO and UDCO techniques. Measurements were obtained from each foal at baseline and during low, intermediate, and high CO states. Measurements were converted to cardiac index (cardiac index=CO/body weight) values for statistical analysis. Agreement between the 2 methods was determined using Bland and Altman analysis and concordance correlation coefficients.
Measurements and Main Results
–
LiDCO determinations of CO ranged between 4.0 and 14.0 L/min resulting in cardiac index ranging between 75.5 and 310 mL/kg/min. There was no significant effect of blood pressure variation on bias or relative bias (P=0.62 and 0.93, respectively). The mean bias and relative bias of UDCO (±SD) compared with LiDCO were −20.1±39.2 mL/kg/min and −7.7±23.4%, respectively. Concordance correlation coefficient between LiDCO and UDCO was 0.833.
Conclusions
–
When compared with LiDCO, the UDCO technique has acceptable clinical utility for measuring CO in healthy anesthetized newborn foals. |
---|---|