Loading…
Blood pressure in children aged 4–8 years: comparison of Omron HEM 711 and sphygmomanometer blood pressure measurements
Objective:To collect normal data on blood pressure (BP) in healthy children aged 4–8 and to compare measurements of BP made in the same subjects with a sphygmomanometer and a portable automated oscillometric BP monitor (Omron HEM 711 with child cuff).Methods:Cross-sectional observational study of 76...
Saved in:
Published in: | Archives of disease in childhood 2009-12, Vol.94 (12), p.955-958 |
---|---|
Main Authors: | , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Objective:To collect normal data on blood pressure (BP) in healthy children aged 4–8 and to compare measurements of BP made in the same subjects with a sphygmomanometer and a portable automated oscillometric BP monitor (Omron HEM 711 with child cuff).Methods:Cross-sectional observational study of 764 children. BP measurements were made at school, using both a sphygmomanometer and an Omron HEM 711. Immediately after the BP measurement children were asked to state which device they preferred (if any).Results:Children had no preference for whether the sphygmomanometer or the Omron was used. Bland-Altman plots showed a lack of consistency between the two methods of BP measurement. With systolic BP there was a trend for the Omron to underestimate when low and overestimate when high.Conclusions:Children were equally distributed in their preference for BP device. There was a wide variation between the two methods of BP measurement, which suggests that comparison of automated BP measurements with normative data obtained by sphygmomanometer is not valid. |
---|---|
ISSN: | 0003-9888 1468-2044 |
DOI: | 10.1136/adc.2008.137059 |