Loading…
Measurements of illuminance in simulated daylight photodynamic therapy
Background Simulated daylight photodynamic therapy (SDL‐PDT) is a new treatment alternative for actinic keratosis. The aim of this study was to show how the illuminance that reaches the target skin area during SDL‐PDT depends on the spatial positioning of the patient. Methods In this technical valid...
Saved in:
Published in: | Photodermatology, photoimmunology & photomedicine photoimmunology & photomedicine, 2022-11, Vol.38 (6), p.564-570 |
---|---|
Main Authors: | , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Background
Simulated daylight photodynamic therapy (SDL‐PDT) is a new treatment alternative for actinic keratosis. The aim of this study was to show how the illuminance that reaches the target skin area during SDL‐PDT depends on the spatial positioning of the patient.
Methods
In this technical validation study, illuminance from the SDL‐PDT system IndoorLux© was measured at different angles, directions, and distances from the light sources corresponding to potential target skin areas. Using two different photometers, data from 63 measuring points at seven specific distances from the ceiling were collected at 0°, 45°, and 90° angles, respectively. Illuminance levels ≥12,000 lux were regarded as adequate. Hotspots were defined as adequate measurements in all directions at a specific measuring point at distances of 1.3, 1.5, and 1.8 m from the light sources (i.e., the most common patient treatment positions).
Results
Adequate illuminance levels were more common with photometer 1 (73%) than photometer 2 (57%). Almost all illuminance levels were adequate at a 0° angle with both photometers. Adequate illuminance levels were observed at 82–93% of the measuring points at a 45° angle and 22–47% at a 90° angle. Hotspots were registered with both photometers at all measuring points at 0°; 59–79% of the measuring points at 45°; and 0–21% at 90°.
Conclusion
Patient positioning is important during SDL‐PDT. Adequate illuminance is achieved if target skin areas are positioned at 0°–45° angles relative to the light sources, but not at 90° angles. |
---|---|
ISSN: | 0905-4383 1600-0781 |
DOI: | 10.1111/phpp.12792 |